Specifications
Instruction 0024-9460 2
Fyrite
®
INSIGHT
CAUTIONS!
• Whentheinstrumentisusedinaninefcientfurnaceorboiler
application where there is a high emission of soot, the probe
sampleltermaybecomeclogged.Checkthelterbeforeevery
usetoconrmthatitiscleanforuseorreplacewithanewlter.
Topreventthisfromoccurring,asmoketestshouldbeperformed
before operating under such conditions to ensure that the furnace
is burning at a level appropriate for the use of this instrument.
• UseonlyBacharachACAdapter(P/N0024-1254).Failuretodo
so could result in damage to the unit and void the warranty.
• Do not store instrument or its sensors with solvents or products
that contain solvents.
1.2 Fyrite
®
INSIGHT Features & Benets
AUTO/MANUAL ZERO - In automatic mode, the instrument automati-
cally zeroes all sensing channels on ambient air when it is powered on. If
a particular sensor is in error during warm-up, the instrument automati-
cally displays the error and continues to operate with the sensor in er-
ror. However, all information dependent on the sensor in error will not
be presented. When the CO channel is set to manual zero, the analyzer
does not zero the CO sensor to ambient conditions during start up. In this
mode, the “fresh air zero” established during manual mode setup is stored
in memory and used during the measurement of CO.
TEMPERATURE AND PRESSURE UNITS - Temperatures can be
displayed in Centigrade or Fahrenheit. Pressures can be displayed in mil-
liBar (mB), Pascal (Pa), hectoPascal (hPa), Millimeters of Water Column
(mmwc), or Inches of Water Column (inwc).
LANGUAGE - The display options are English, French or Spanish.
MEMORY - The Fyrite
®
INSIGHT has the ability to store, recall to dis-
play, and print a minimum of 100 sets of time and date coded combustion,
differential pressure, or differential temperature test records.
DOWNLOAD - Test records can be downloaded to a personal computer via
the USB port. Saved records are transferred using PC based software and
will be automatically loaded into an Excel spreadsheet. PC based software
will also have the capability to update the instrument time and date and
upload custom fuels, username, customer logo, and future code updates.
The INSIGHT can also be set up from PC software.