Specifications

2
WARNING: This deviceis not to be
used in any application that is beyond its
intended purpose or beyond the scope of its
specifications. For details on appropriate
use, refer to the rest of this manual. Before
risking equipment damage or personal
injury, contact Bacharach if you are unsure
of the validity of a particular combustion
analyzer application.
How the H-10PM Works
The sensor in the H-10PM uses positive
ion emission technology, commonly known
as a heated diode. It is very sensitive to only
halon substances (refrigerants) making this
product highly resistive to false alarming,
while retaining superior sensitivity for pin-
pointing the most difficult to find refriger-
ant leaks.
A pump inside the unit draws air through
the sensor. Any presence of halogen gases
(such as refrigerants) causes an ionized
current to flow that sounds a speaker
and illuminates a neon light in the probe.
Sensitivity to pinpoint both large and small
leaks can be controlled by setting the three
position Leak Size switch.
Panel Operating Controls
Charger
Plug
Battery
Charge
Light
Calibration
Reference
Bottle
Sensor
Leak Size
Switch
Manual Balance
Control
Heater
Adjustment (turn
clockwise
to increase heat)
Power
Switch
Low Battery
Indicator
Heater Adjust
LEDs
Mode Switch
MANUAL /
AUTO