4.3. Relays
Connect leads to terminal block for Common (COM) and N/O and/or N/C
connections as required. Note that relays are rated as 10A @
120/230 VAC.
Relay Wiring
Type Two-Alarm Units One-Alarm Units
1-2 Channel
Units
CN5: Low-Level Alarm
CN4: High-Level Alarm
CN4
4-6 Channel
Units
CN10: Low-Level Alarm
CN9: High-Level Alarm
CN12: Fault Relay
CN10
NOT E: N/O (normally open) and N/C (norm ally closed)
refer to
contact status in standby mode. On a two-alarm system, a
high-level alarm condition on any sensor will override a low-
level alarm condition on another sensor.
NOT E: On 4- to 6-channel, two-alarm units, the high-level
relay may be set for normal or Fail-Safe operation by setting
jumper JP1 on the cont rol unit’s printed circuit board.
4.4. Power Connection
Use 3-wire, 20 AWG wi re for 230V systems or 3-wire, 18 AWG wire for
120V systems. Connect the m ain power s upply t o terminal block CN3
(on 1 & 2 channel systems), or fused t erminal block mounted on base of
control unit (4 - 6 channel systems). Ensure that ground connections to
the lid and base of the enclosure are maintained.
NOT E: Connection to the main power supply must be made
via an approved, readily-accessible, switched and fused plug
and socket (or as per local wiring regulations) which should be
within 10 feet (3 meters) of the cont rol unit.