Specifications
M GD-100 M anual
36 6109-9000 Rev 3
Step Electrical Re set of One-Alarm Systems
2
Connect a voltmeter between Pins 4 (-Ve) and 2 (+Ve) of the
sensor terminal connector block for each channel in turn
(CN1, CN2, etc.) and adjust the corresponding calibration pot
(P1, P2, etc.) to the sensor standby voltage (SSV) value as
per calibration label on side of enclosure. (For IR sensors,
set SSV to 2V.) See Figure 13 and Figure 15.
3
The alarm threshol d volt age (A TV) is the voltage at which the
alarm and relay acti vat e at a given gas concentration. This
voltage is normally set to 3.0V at the factory and is adjusted
using:
• pot “P3” for 1- or 2-c hannel units (see Figure 1)
• pot “P7” for 4- or 6-c hannel units (see Figure 2)
Measure bet ween pi ns 4 (-Ve) and 1 (+Ve) on the Cal
header. See Figure 13 and Figure 15.
4
For the al arm set point, connect the DC voltmeter between
0V (TP5) and High (TP1) (see on page 10). Adjust P7 to the
alarm set poi nt as per the rating label (normally 1.2V ).
5
Carry out a bump test to ensure the sensor is functioni ng
correctly.
6
If the sensor does not go into alarm, carry out a gas
calibration.
7
Finally, return jumper JP1of the controller to its original
position.
6.3. Electrical Reset of Two-Alarm Systems
Reset, if necessary, the Standby and low /high Alarm Threshold Voltages
to the factory settings as shown on the calibration label. This is
performed on the sensor P CB.
Electrical reset information is listed on the label on the side of the
enclosure and is unique to that sensor.
Tools required:
• A voltmeter (crocodile clips recommended)
• Factory set point electric values (as shown on the rating label)
• Screwdri ver (depending on enclosure).










