Specifications

M GD-100 M anual
40 6109-9000 Rev 3
Step Bump Testing Using Ga s Ampoule s
9
Remove the beaker from the sensor. Carefully remove any
ampoule remains from the gas detector and beaker.
6.5. Calibration Overview
To comply with t he requirements of EN378 and t he E uropean F-GAS
regulation, sensors must be tested annually. However, local regulations
may specify the nature and frequency of this test.
CAUTION: Check local regulations on c alibration or t esting
requirements.
6.6. Calibration Options
There are two available calibration options:
Exchanging the sensor board for a new, pre-calibrated one and
send old one to B acharach for rec alibration (available for two-
alarm units and 1- and two-alarm IR units)
On-site gas calibration.
These are explained in the next two sections.
6.7. Sensor Board Exchange
There are a number of advantages to sensor board exchange. It is
simpler and quicker than gas calibration.
NOT E: S ensor board exchange is available for t wo-alarm
semiconductor (SC) units and one- and two-alarm
infrared
(IR) units.
NOT E: Bacharach recommends exchangi ng your sensor
PCB for a new pre-calibrated one every 3 years.
Tools required:
A pre-cali brated sensor board
A voltmeter (crocodile clips recommended)