Specifications
Instruction 0019-9321
Comfort Chek 100/200
4
2.3 Pump Operation (Model 200)
With the instrument already switched ON, momentarily pressing the
button will start and stop the internal pump. The optional Extended
Probe Assembly with in-line fi lter (P/N 0019-3310) can be connected to
the gas inlet fi tting (Figure 1, Item A) on the top of the instrument for
drawing in gas samples from hard-to-reach areas. If desired, a section
of
1
/
8
" I.D. tubing can be connected to the gas outlet fi tting (Figure 1,
Item B) to exhaust the gas being sampled to an outside area.
Note that the combined sampling hose and probe length should not
exceed 6 feet (1.8 m). Also note that the internal pump is intended for
use only at normal atmospheric pressure, and is not designed to draw
in gas samples against a vacuum or an obstruction such as a kink in the
sampling hose. If an obstruction or negative-pressure gradient is pres-
ent, then gas will not be drawn into the instrument. Please consult the
factory for applications where longer sampling lengths are required, or
where it is necessary to draw against a vacuum.
2.4 Pump Contamination (Model 200)
Over time the pump can become contaminated, leading to a slower
response time and lower readings. To check for pump contamination,
turn ON the pump and hold your index fi nger over the inlet fi tting. The
pump should normally stall. If the pump, however, continues to run, then
the pump is contaminated and the instrument needs to be returned to
Bacharach for repair.
2.5 Selecting the Measurement to be Displayed
Depending upon which model of the Comfort Chek was purchased, sev-
eral different gases and environmental measurements can be taken as
described in Section 1.1 General.
Important! The temperature and humidity sensor probe (Fig-
ure 1, Item D) must be installed for the instrument to operate
properly. If this probe is removed, inaccurate temperature and
humidity readings will be displayed.
Each measurement is displayed separately on the instrument’s LCD
(Figure 1, Item C). The measurement displayed is selected by pressing
the f
3
button.