Specifications

Instruction 0024-9395
17
Fyrite Pro
4.0 OPERATION
4.1 Performing CO and Draft Testing
4.1.1 Analyzer Turn On and Warm Up
1. Connect the probe and make sure that the analyzer is properly set
up per Sec tion 3.0.
2. Turn ON the analyzer in fresh air by pressing is I/O button and
observe that the following Warm-Up Screens are displayed.
These screens show the analyzers model number, the units soft-
ware revision, and finally the remaining warm-up time. If an O
2
sensor is installed, then its current output level is also displayed.
Note: The warm-up time is 10 seconds for analyzers that do
not contain an oxygen sensor, and have their CO channel
set up for manual zero. The warm-up time is 60 seconds for
analyzers that contain an oxygen sensor, and/or have their
CO channel set up for auto zero.
Tip: An O
2
error will occur when the oxygen sensor’s output
drops to between 80 and 90. Consider replacing the oxygen
sensor when its output level drops below 100.
3. Wait for the analyzer to count down its warm-up period; after
which, either the CO Screen (Model 100), the CO & Draft Screen
(Model 105), or the Fuel Screen (Model 110/120/125) is displayed.
Model 100 Model 105 Model 110/120/125
If, however, any errors were detected during warm-up, the Sensor
Status Screen is displayed. If this occurs, refer to Section 4.13.13 for
information on how to correct the error.