Specifications

Instruction 24-9395
42
Fyrite Pro
5.2 Ambient Temperature Calibration
(Model 110/120/125)
Material Required: Calibrated Thermometer
Procedure:
1. Enter the Calibration Mode as described in Section 5.1. Observe
that “AMBIENT” will appear at the top of the display; if not, re-
peatedly press ENTER until “AMBIENT” is displayed.
2. Use the and buttons to set the displayed value to match the
reading of a calibrated thermometer at room temperature.
3. Press ENTER to store the displayed value and move to the next
calibration screen, or hold down ENTER for 2 seconds to store the
displayed value and enter the Run Mode, or press the I/O button to
exit the Calibration Mode and turn OFF the analyzer without sav-
ing the changes.
5.3 Combustion Air Thermocouple
Calibration (Model 125)
Material Required: Thermocouple Simulator
Procedure:
1. Attach the thermocouple simulator to the Combustion Air Thermo-
couple connector on the analyzer (T-AIR, see Figure 3-2).
2. Enter the Calibration Mode as described in Section 5.1. Then re-
peatedly press the ENTER button until “TA” appears at the top of
the display.