Specifications

Instruction 24-9395
43
Fyrite Pro
3. Set the simulator to 32.0 °F (0 °C); then use the and buttons to
set the displayed value to match the simulator’s value.
4. Set the simulator to 212.0 °F (100 °C); then use the and but-
tons to adjust the displayed value to match simulator’s value.
5. Repeat Steps 3 and 4 as necessary until the unit is reading correctly
at both temperatures.
6. Press ENTER to store these calibration values and move to the next
calibration screen, or hold down ENTER for 2 seconds to store these
calibration values and enter the Run Mode, or press the I/O button
to exit the Calibration Mode and turn OFF the analyzer without
saving the changes.
7. Remove thermocouple simulator from analyzer.
5.4 Stack Thermocouple Calibration
(Model 110/120/125)
Material Required: Thermocouple Simulator
Procedure:
1. Attach the thermocouple simulator to the Stack Thermocouple con-
nector on the analyzer (T-STACK, see Figure 3-2).
2. Enter the Calibration Mode as described in Section 5.1. Then re-
peatedly press the ENTER button until “STACK” appears at the top
of the display.
3. Set the simulator to 32 °F (0 °C); then use the and buttons to
set the displayed value to match the simulator’s value.
4. Set the simulator to 572 °F (300 °C); then use the and buttons
to adjust the displayed value to match simulator’s value.
5. Repeat Steps 3 and 4 as necessary until the unit is reading correctly
at both temperatures.