Specifications
MODEL 300 COMBUSTION ANALYZER
3-10 Instruction 24-9223
2. If a recalibration cycle is started, allow the cycle to finish before
proceeding.
If RUN is displayed, press the RUN key to restart the efficiency test.
If FUEL is displayed, press the FUEL key until the desired fuel is
shown in the left display. Then press the RUN key to restart the
efficiency test.
3. If the “error/warning/overrange” indication reappears after three
attempts have been made to clear it, correct the problem(s) displayed
in Step 1 by following the suggested remedies listed in Table 3-1.
3.5.2 Optional Error, Warning, and Overrange Displays
The warning, failure, and error messages listed in Table 3-2 are displayed
on the auxiliary-display panel when an abnormality is detected within the
NSX portion of the analyzer.
Since the detected values of NO
X
and/or SO
2
are not used in the calcula-
tion of combustion efficiency, the analyzer will continue to operate and
provide valid combustion efficiency test results when any of the messages
listed in Table 3-2 are displayed.