Specifications
MODEL 300 COMBUSTION ANALYZER
5-4 Instruction 24-9223
7. Remove shorting wire from CO and/or SO
2
sensor terminals R and S.
8. Loosen nuts on sensor terminals C, S and R; then connect wires from
analyzer to the sensor as shown in Fig. 5-1. Tighten these connections
with a nutdriver. DO NOT overtighten connections!
9. Replace sensor cover.
10.Calibrate the new Toxic sensor per Section 4.3 (See Table 5-1 for NO
X
stailization time prior to calibration).
Figure 5-2. Toxic Sensor Replacement