Specifications
MODEL 300 COMBUSTION ANALYZER
5-14 Instruction 24-9223
5.7 OPTIONAL SENSOR BIAS BATTERY REPLACEMENT
Four “AA” batteries mounted behind the control panel apply a constant
bias voltage to the NO
X
and/or SO
2
sensor(s). The batteries should last a
minimum of 6 months, with a typical life expectancy of 9 months to more
than 1 year.
Replace the batteries when the LOW BATT message appears on the
auxiliary display panel. You have approximately 1 month of operating
time left after the first LOW BATT message is displayed before the
batteries go completely dead, as indicated by the BATT FAIL message.
Replace the batteries as follows:
Equipment and Materials Needed:
• Medium Flat-Blade Screwdriver • Four “AA” Alkaline Batteries
Procedure:
1. Remove control panel per Section 4.1 in this manual.
2. Tilt the control panel upwards until the four “AA” batteries become
accessible (Fig. 5-7).
3. Replace batteries with four fresh “AA” alkaline cells in the proper
polarity.
If batteries are installed incorrectly, the BATT FAIL message appears
on the auxiliary-display panel after analyzer is powered on and run
through its calibration cycle.
The SO
2
sensor stabilizes within 15 minutes after the application of
bias voltage.
Depending on how long the NO
X
sensor was without bias voltage, the
time required for this sensor to stabilize varies from less than a
minute to several days. Typical stabilization times are shown below:










