Service manual

Single Gas Instrument Service Manual
Page 4 Bacharach, Inc. Instruction 0019-9166
1.0 SERVICE MANUAL CONTENTS
This manual contains the following service information for the . . .
DIOXOR II (Part No. 0019-7038 & 0019-7043)
MONOXOR II (Part No. 0019-7034 & 0019-7039)
MONOXOR II H (Part No. 0019-8015)
NONOXOR II (Part No. 0019-7036 & 0019-7042).
NOTE: The 0019-7039, 0019-7042, and 0019-7043 versions of these
instruments do not contain a probe, all non-probe related functions
remain the same in this instruction.
Instrument maintenance, which includes information on how to:
Open and close the case
Clean the Fan Assembly and chamber assembly area
Replace the sensor
Replace the Printed Circuit Board (P.C.B.)
Service the gas-sensing chamber assembly (chamber assembly)
Sensor calibration
Troubleshooting chart
Replacement parts list
GENERAL NOTE
If excessive torque is applied to the instrument's
fasteners, poor performance or damage to components
may occur.
The torque speci cations are as follows:
Screws Torque
Sensor Mounting 4 - 8 oz. in.
Motor Mounting 6-10 oz. in.
Chamber Sealing 16-20 oz. in.
Rear Cover/Case 18-24 oz. in.
Fan Assembly 32-40 oz. in.