Specifications

Instruction 0019-9326
8
Monoxor III
3.4 Setup Mode
The analyzer is preset at the factory for the parameters shown below, but
can be changed as described in their associated sections.
Function Parameter To Change
Temperature Unit °F Section 3.4.2
CO Channel Auto Zero Section 3.4.3
Time Not Set Section 3.4.4
Date Not Set Section 3.4.5
Year Not Set Section 3.4.6
Printer IrDA Section 3.4.7
3.4.1 Entering Setup
1. With the analyzer turned OFF, press and hold down the HOLD
button.
2. Press and release the I/O button.
3. Release the HOLD button.
4. The analyzer is now in its Setup Mode. Refer to Sections 3.4.2 thru
3.4.7 for information on how to set the analyzer’s various parameters.
3.4.2 Temperature Units Setup
The Temperature Units Setup Screen is labeled “Unit”.
Note: The analyzer uses an internal temperature sensor
to maintain CO sensor accuracy under varying ambient
temperature conditions. This sensors temperature reading is
displayed only during calibration as described in Sec-
tion 5.2.