Specifications

Instruction 0019-9326
22
Monoxor III
5.4 Water Trap / Filter Maintenance
The Water Trap / Filter Assembly removes water condensate from the gas
sample, and prevents soot from contaminating the internal components of
the analyzer. See Figure 5-2.
Drain the water condensate after every test and during each test if the
water level nears the tip of the riser tube in the water trap side.
Procedure:
1. Pull apart the water trap portion of the water trap/lter assembly
using a twisting motion.
2. Pour out all water condensate; then reassemble the trap.
Replace the Filter Element when dirty.
Material Required: Filter Element, (P/N 0007-1644)
Procedure:
1. Pull apart the lter chamber of the water trap/lter assembly using
a twisting motion.
2. Remove and discard the old lter.
3. Install a new lter element (P/N 0007-1644)
4. Reassemble the lter chamber and water trap/lter assembly.
IMPORTANT: Ensure that the surfaces of the water trap/lter
assembly make contact.
Figure 5-2. Water Trap / Filter Assembly
Probe
Filter
Water Trap
Hose
(To Monoxor III)
Water Trap/Filter Assembly