Specifications

Technical Characteristics
PCA
Normal Operating Conditions:
Temperature:
Analyzer
...................
32 to
104 ºF (0 to 40 ºC)
Probe Tip
..................
1472 ºF
(800 ºC) Max.
Humidity:
Analyzer
...................
15 to
90% Relative Humidity, Non-Condensing
Air Pressure:
Analyzer
...................
Atmospheric
Probe .........................
10"
H
2
O (25 mb) draft max at probe
tip
Performance:
Accuracy:
Oxygen*
.................... ±0.3% O
2
Carbon
Monoxide
.....
±5% of reading or ±10
ppm, whichever is greater
between 0 – 2000 ppm, and ±10% of reading
between 2001 – 4000 ppm.
Nitric Oxide
..............
±5% of reading or ±5 ppm, whichever is greater
Flue Gas Temp.
........
±4 ºF between 32 and 255 ºF
(±2 ºC between 0 and 124 ºC)
±6 ºF between 256 and 480 ºF
(±3 ºC between 125 and 249 ºC)
±8 ºF between 481 and 752 ºF
(±4 °C between 250 and 400 °C)
Ambient
Temp.
........
±2 ºF between 32 and 212 ºF
(±1 °C between 0 and 100 °C)
Pressure
...................
±2% of reading or ±0.02 inches of Water Column
(±0.05 mb), whichever is greater
System Flow Rate:
With
probe ................ 200
cc/min minimum
Front Panel
Controls:
Seven embossed pushbutton switches with tactile feedback (refer to
Section 4.3)
Display:
20 character by 4 line alphanumeric LCD panel with a green backlight.
*
Accuracy referenced in practical flue gas concentrations
(mixtures of O
2,
CO
2
and N
2
)
2-2
Instruction 0024-9219