Specifications
PCA
®
3
21
3.2.2 Using the AC Power Adapter
The AC power adapter is capable of powering the analyzer on a continuous
basis. The adapter plugs into an appropriate 100-240 VAC, 50/60 Hz wall
outlet, and produces an output of +6 VDC. The adapter’s output connector
plugs into the analyzer’s POWER jack located on the botton of the unit
(Figure 3-2).
3.3 Connecting the Probe & Hose Assembly
To attach the probe and hose assembly to the analyzer (Figure 3-2):
1. Push the gas-sample hose connector, the larger of the two
connectors (giving a slight twist), onto the analyzer’s GAS fi tting.
2. Push the draft-hose connector, the smaller connector (giving a
slight twist), onto the analyzers +∆P fi tting.
3. Push the stack-gas thermocouple connector into the T-STACK jack
(connector fi ts in only one way).
4. Push the optional primary/ambient air thermocouple into the
T-AIR jack (connector fi ts in only one way).
IMPORTANT: To assure the accurate calculation of
combustion effi ciency, the optional primary/ambient air
thermocouple must be used when the burner’s primary-air
temperature is not the same as the room temperature.
5. Inspect all hoses for cracks. If any hose is found to be defective,
replace the entire probe and hose assembly. Check that the water
trap is empty, and the fi lter is not dirty or saturated with water.