Specifications
PCA
®
3
38
3.19 Cal Reminder Period
The analyzer can be set to indicate a calibration reminder during the 60
second warm-up period. Calibration reminders can be preset to occur
never, 6, 8, 10, 12, or 15 months after the last calibration. When the
preset period is exceeded the instrument will display the reminder, and
how long since the sensors were last calibrated. The reminder will be
displayed at the end of the 60 second warm-up period. If a calibration
reminder is displayed the operator can press the RUN/HOLD key to
move to the Run/Hold screen for normal operation. Regular calibration
periods of 6 months to a year for all gas sensors (except Oxygen) are
recommended.
Set the calibration reminder period as follows:
1. Display the MAIN MENU by
pressing the MENU (F2) button. If
necessary, press ESC until MENU
appears above F2.
2. Use the ▲▼ buttons to highlight
SETUP, and then press ENT to
display the SETUP MENU.
3. Use the ▲▼ buttons to highlight
CAL REMINDER PERIOD, and then
press ENT to display CAL Reminder
Period Menu.
4. Use the ▲▼ buttons to highlight the
desired CAL reminder period, and
then press the ENT key to activate
the selection and re-display the Setup
Menu.
NOTE: The date and time settings
must be correct to get accurate cal
reminders.