Specifications
PCA
®
3
45
4.4 Performing a Combustion Test
Ensure that the following has been completed, and then proceed with the
combustion test as described below:
• Turn ON analyzer and allow it to
warm up (Section 4.2).
• Select fuel being burned (Section
3.5).
• Inset probe into stack (Section 4.3).
• If necessary, insert optional primary
air thermocouple into combustion-
air stream of burners that use an
outside source of combustion air.
1. Press the RUN/HOLD button to start the test. You should hear the
pump start running and see the word RUN appear at the top of
the Combustion Test screen.
Sensor Indicators: The following indicators appear in the
sensor’s data fi eld depending on certain conditions:
(* * *) Sensor that is not calibrated or installed.
(XXX) Sensor overrange
(- - -) The calculated data cannot be displayed
because the measured data necessary to make the
calculation is out of range (i.e., oxygen level above
16%).
2. Use the up and down arrow keys to scroll to the T-STK reading.
Loosen the thumbscrew on probe stop and move probe in and
out of the stack until the stack’s core temperature (hot spot) is
located as indicated by the highest T-STK reading; then tighten
thumbscrew to prevent further probe movement. Locating the
highest stack temperature is very important for accurate effi ciency
calculations.
3. You can now begin burner-service procedures. The analyzer
readings will change quickly to show changes in burner
performance.
CAUTION: Position the Water Trap with its gas-fl ow arrow
pointing upward. Do not let water condensate go above the
tip of the riser tube. The sensors could be damaged if water
would enter the analyzer. Empty the Water Trap after every
combustion test (refer to Section 4.10).