4.7 Making a Draft / Pressure Measurement
The difference in pressure (P) between two areas can be measured by
using the analyzer’s two pressure ports and the PRESSURE screen. By
using the –P port as the reference, the pressure applied to the +P port
will be displayed on the PRESSURE screen as the differential pressure
between the two ports.
Perform a draft / pressure measurement as follows:
1. Turn ON the analyzer and allow it to
complete its warm-up cycle ( Section
4.2).
TIP: The pressure units of measure is
selected per Section 3.7.
2. Display the MAIN MENU by
pressing the MENU (F2) button. If
necessary, press ESC until MENU
appears above F2.
3. Use the ▲▼ buttons to highlight
PRESSURE, and then press ENT to
display the PRESSURE screen.
4. Before taking a measurement, the
pressure sensor may need to be
re-zeroed if it is not already displaying
zero with both pressure ports open to
the atmosphere. If necessary, zero the
pressure sensor as follows:
• Press the ZERO (F2) button.
Disconnect any hoses connected
to the +P and –P ports, and
then press ENT to zero the
pressure sensor.
Reconnect any hoses. When
measuring draft, simply leave the
P port open to the atmosphere
and connect the probe’s draft
hose to the +P port (see Figure
4-2).
5. Do one of the following to measure draft or differential pressure: