Specifications

PCA
®
3
50
To measure draft, simply insert the probe into the
stack and observe the draft reading on the PRESSURE
screen.
To measure differential
pressure, connect two sampling
hoses to the +P and –P ports,
and place the open end of each
hose into the areas being
measured. The differential
pressure between the two areas
is now displayed on the
PRESSURE screen. If the
pressure at the +P port is
higher than the –P port, then the pressure reading
will be positive. If it is lower, then the reading will be
negative.
4.8 Saving Test Data
Up to 500 individual sets (“snap shots”) of combustion-test, pressure, or
temperature data can be saved in memory, which can later be recalled for
viewing from the Memory Directory (Section 4.14).
1. First display the screen that contains
the data to be saved. In the example
to the right, all data associated with
the Combustion Test RUN screen
will be saved.
2. Press the SAVE (F3) button to save
the test data in the next available
memory location.
NOTE: When memory is full, the next
reading will not be saved until space
is made available by clearing the
data. (See 4.14.3: Clearing Memory)
POWER
P
T-STACK T-AIR
GAS
P (Pressure
Reference Port)
+P (Pressure
Measurement Port)
Figure 4-2. Draft/Pressure Measurement Hose Connections