4.11 Turning OFF the Analyzer & Purging
Turn OFF the analyzer by pressing the I/O button for at least 2 seconds,
or until two beeps are heard. The unit will count down 5 seconds before
shutting down, giving the operator an opportunity to keep the analyzer
turned ON by pressing the RUN/HOLD button.
If the PCA
®
3 was not purged with fresh air as described in Section 4.9,
then the analyzer may remain ON with its pump running and display the
message “PURGING SENSORS” as the result of combustion gases still
being present inside the analyzer. At this time the operator should ensure
that the probe is removed from the stack, allowing the analyzer to purge
itself with fresh air. The 5-second shutdown sequence will not begin until
the gas levels inside the analyzer drop below predetermined levels:
TIP: Although not recommended, the purging process can
be bypassed by pressing the I/O button a second time.
4.12 Low Battery Alarm
When the batteries are nearly depleted, an empty battery icon appears in
the upper-right corner of the display, and a short beep is sounded every
10 seconds.
After a low battery alarm occurs, the analyzer will continue to operate for
only a few minutes. The amount of operating time that remains depends
on many factors (e.g., pump and keypad backlight being ON or OFF, and
the type and condition of the batteries).
4.13 Data Logging
When the logging function is activated, up to 500 combustion test records
will be automatically stored in memory at a preset interval (1, 5, 10, 15,
30 seconds, 1, 2, 5, 10 minutes) over a predetermined duration (5, 10, 15,
30 minutes, 1, 2, 5, 10, 24, 48 hours).
The maximum duration that data can be collected is determined by
the interval. For example, if the interval is set to 10 seconds, then the
maximum selectable length of time in which data can be collected to ll
500 memory locations would be 1 hour (500 x 10 seconds = 5000 seconds
or 83 minutes). If the operator chooses a duration that is longer than
possible for the interval chosen, then the analyzer automatically selects
the highest duration possible for the selected interval.