4.13.3 Starting the Data Logging Process
After turning ON data logging and setting
the interval and duration, the analyzer will
pause at the LOGGING SUMMARY screen,
where the currently selected interval and
duration time periods are displayed.
Press ENT to start the combustion test and
logging process. At this time the Combustion
Test LOG screen will appear, indicating that
the analyzer is now performing a combustion
test and the data is being stored in memory.
Note the following:
• At the top of the screen, the memory
location where the current log entry
is being saved is displayed.
• If there were previously stored log
entries, each new logging session
will start to store data in the next
available memory location.
• At no time will new data overwrite
old data.
4.13.4 Ending the Data Logging Process
Data logging will stop and the pump will turn OFF after the prescribed
duration, or after all 500 memory locations are fi lled.
To exit the Combustion Test LOG screen, press the RUN/HOLD button
twice to display the Combustion Test HOLD screen.
Note the following:
• To end the logging process at any time, press the ESC button.
• If the logging process was stopped by pressing ESC, it cannot be
resumed without fi rst turning logging back ON per Section 4.13.1.