Specifications
PCA
®
3
70
5.3 B-Smart
®
Sensor Replacement &
Calibration
To replace a B-Smart
®
sensor and calibrate, do the following:
1. Enter the CALIBRATION MENU per Section 5.2.
2. Use the ▲▼ buttons to select B-SMART
®
. Press ENTER to display
the B-Smart code screen
3. Use the ▲▼ buttons to enter the 10 or 14 digit alphanumeric code
supplied with the B-Smart
®
sensor. Use the ◄► buttons to move
the cursor across the screen. Press ENTER.
NOTE: There are two B-Smart
®
sensor code lengths. 10
digits for CO
High
, NO, NO
2
, and SO
2
, and 14 digits for
CO
Low
.
NOTE: If the correct code was entered, the analyzer
accepts it and returns to the CALIBRATION MENU.
If an incorrect code was entered, the screen will display
“Invalid Code”. Check to make sure the correct code has
been entered. If the problem persist, contact your nearest
Bacharach Service Center.
TIP: To save time B-Smart
®
codes can be entered using the
Fyrite User Software.
Bacharach also offers a convenient Exchange Program that allows the
customer to regularly send in old sensors and, as scheduled, receive
calibrated replacements. Contact Bacharach customer service for more
details about the Exchange Program.