5.4 Pressure Sensor Calibration
This procedure calibrates the pressure sensor to a known pressure value.
Material Required:
• Bellows
• Manometer
• Range: ±8 in. of water column (±20 mb)
• Accuracy: ±0.01 in. of water column (±0.025 mb)
Procedure:
NOTE: The unit-of-measure for pressure is selected per
Section 3.7. In the following procedure inwc is selected, but
note that any unit-of-measure can be used for calibration
purposes.
1. Assemble the pressure sensor calibration equipment as shown in
Figure 5-1, but DO NOT connect the analyzer to the calibration
equipment at this time.
2. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen per Section 5.2.
3. Use the ▲▼ buttons to highlight
Pressure, and then press ENT to
display the CALIBRATE PRESSURE
screen.
“Measured” is the pressure value
currently being detected by the pressure
sensor, while “Applied” is a known
value of pressure that will be applied
for calibration purposes.
4. With both the –∆P and +∆P ports
open to the atmosphere, observe
that the current Measured pressure
reading should be 0 ±0.01 inwc. If
necessary, zero the pressure sensor
per Section 4.8, and repeating Steps 2
thru 4.
5. Connect the hose from the manometer
to the +∆P port and apply a negative
pressure to this port by adjusting the
bellows for a manometer reading of –4.00.