Specifications

PCA
®
3
80
5. Adjust regulator for a owmeter indication of approximately
2 SCFH. Wait until the Measured reading stabilizes (approximate-
ly 3 minutes), and then press ENT to calibrate the SO
2
Measured
value to that of the Applied value. The message “Good Calibration”
should brie y appear followed by the CALIBRATION LIST screen
being re-displayed.
If the sensor’s output is low, but still usable, then the
message “Good Calibration WARNING Low Sensor” will
appear. The sensor will now be marked as being Low in the
DIAGNOSTICS screen.
If the sensor’s output is too low to be usable, then the
message “Bad Calibration Sensor End of Life, Entry Not
Saved” will appear followed by the CALIBRATION LIST
screen being re-displayed. The sensor will now be marked
as being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove gas cylinder.
5.9 NO Sensor Calibration
This procedure spans the optional nitric oxide sensor to a known gas level.
Material Required:
Calibration Kit, P/N 0024-7059
Gas Cylinder: 250 ppm NO in Nitrogen, P/N 0024-1156
Procedure:
1. If not already done, turn ON
the analyzer and display the
CALIBRATION LIST screen per
Section 5.2.
2. Use the ▲▼ buttons to highlight NO,
and then press ENT to display the
CALIBRATE NO screen.
“Measured” is the current NO reading,
while “Applied” is a known NO level
that will be applied for calibration purposes.
3. Attach a 250 ppm NO cylinder to the regulator of the calibration
xture shown in Figure 5-2 on Page 85.
4. Use the ▲▼ and ◄► buttons to enter an Applied value that
exactly equals the concentration stamped on the NO cylinder.