6.5 Thermocouple Replacement
Using the appropriate thermocouple replacement kit listed below, replace
the probe’s thermocouple as follows:
Thermocouple Replacement Kits:
Part Number Replaces Thermocouple in a
Probe with a Tube Length of...
0024-8413 6 inches
0024-8414 12 inches
0024-8415 24 inches
0024-8416 36 inches
Each Kit contains a thermocouple assembly, two O-rings, and two wire
splice connectors.
Tools Required:
Small Flat Blade Screwdriver
Wire Cutter
Wire Stripper
Slip Joint Pliers
Procedure:
1. Gain access to the thermocouple connections by rst removing
three screws from probe handle, and then separating the two
handle pieces.
2. Cut wires attached to old crimp connectors, leaving behind as
much of the probe’s thermocouple-connector wire as possible.
Probe Handle
Screws