Specifications

PCA
®
3
95
3. Pull old thermocouple from probe body and discard.
4. The new thermocouple has been coiled for shipping purposes.
Straighten the thermocouple using your thumb and index nger.
5. If not already done, install supplied O-Rings onto thermocouple.
6. Insert thermocouple into probe body until it “bottoms out.”
7. Strip 1/4 inch of insulation from each of the probe’s thermocouple
connector wires.
IMPORTANT: In Step 8, the thermocouple wires must rst
be twisted together and then crimped.
8. Twist both red thermocouple wires together; insert them into the
supplied wire-splice connector; and then crimp the connector using
a pair of pliers. Repeat this step for the yellow thermocouple wires.
9. Reassemble the probe handle, being careful not to pinch the
thermocouple wires between the handle pieces. In addition, ensure
that the end of the thermocouple is in front of the rib molded into
the bottom handle piece; otherwise, the handle pieces will not t
tightly together.
O-Rings
Thermocouple Assembly
Ensure end of Thermocouple is
in front of plastic rib in handle
Wire-Splice Connectors
Red and Yellow
Thermocouple Wires
Cut away view of Thermocouple
inserted into Probe Body
When reassembling probe
handle, ensure that the thermo-
couple wires do not become
pinched around screw hole or
edge of handle
Figure 6-8. Thermocouple Installation and Wiring