Specifications

PCA
®
3
97
7 Troubleshooting
7.1 Analyzer Repair
It is recommended that eld repair of the PCA
®
3 be limited to:
Checks of printed circuit board connectors
Replacing the probe assembly
• Replacing the lter element in the water trap / lter assembly
• Replacing sensors
Information on how to perform these repairs is provided in Section 6.
All other repairs should be performed by an authorized Bacharach Service
Center (refer to Section 8.3). Any repairs performed by an unauthor-
ized service organization will void the analyzer’s warranty and release
Bacharach, Inc. of any implied or written product liability.
7.2 Error Symbols
Error symbols are shown in the data elds of the Combustion Test screens
and on the printout to indicate the following:
- - - Three hyphens or dashes indicates that the calculated data cannot
be displayed because the measured data necessary to make the calculation
is out of range. For calculations to occur, the oxygen level must be below
16% and the stack temperature must be below 2,000 °F (1,093 °C).
“* * *” Three stars appear in the data eld of sensors that are not
installed, defective, or found to be in error during warm-up, and also in
the data elds of that sensor’s related calculated values. For example, if
the data eld of either the NO or NO
2
sensor contains three stars, then the
NOx data eld will also contain three stars.
“X X X” Indicates sensor over range.