Manual

SPECIFICATIONS
Materials of Construction
Body 303 stainless steel.
Rotor CD4MCU stainless steel.
Bearings Two (2) type 440 stainless steel ball bearings.
Rotor Support and Shaft 303 stainless steel.
Operating Parameters
Temperature
Meter –60…350° F (–51…177° C).
RF Pickup –150…325° F (–101…162° C).
Pressure 4000 psi (276 bar) maximum.
Accuracy ±0.5% of reading.
Repeatability ±0.1%.
Calibration Water (NIST traceable calibration).
Corrosion
All Blancett Model 1200 turbine meters are constructed of stainless steel. The operator must
ensure that the operating fluid is compatible with these materials. Incompatible fluids can
cause deterioration of internal components and cause a reduction in meter accuracy.
Pulsation and Vibration
Severe pulsation and mechanical vibration will affect accuracy and shorten the
life of the meter.
Filtration
If small particles are present in the fluid, Blancett recommends installation of a strainer
upstream of the meter (see Table 1 for filtration recommendations).
RF Preamplifier
Input Signal 1 millihenry carrier pickup.
Output Signal 10V peak to peak square wave.
Temperature
Module –20…160° F (–7…71° C).
Pick-up –150…325° F (–101…162° C).
Power 7…30V DC.
Distance Specification
50 ft maximum between pickup and RF preamplifier.
1000 ft maximum between preamplifier and receiving unit.
Electrical Connection Terminal strip.
Housing Epoxy encapsulated module.
Turbine Flow Meter, Model 1200 Turbine Flow Meter
Page 6 November 2013