Data Sheet

Industrial line registers
Type ILR 700, 701, 720, 740, 750(T)
AC_ILR_DB_02_1662 (October 2019)
Data Sheet
FEATURES
Large six-digit LCD display
Display in liters, pints, quarts or gallons
11 digits, non-resettable totalizer, 6-digits resettable
totalizer
Temperature -20 °C to +80 °C (-4 °F to +140 °F)
Replaceable long life battery (not used if external power
connected)
field programmable
9-point linearization (ILR 750, ILR 701)
Scalable pulse output (ILR 750)
programmable analog (4-20mA) output (ILR 750)
Protection class IP65
DESCRIPTION
The electronic register module contains a microprocessor
board powered by a lithium battery. It can be programmed
to measure in liters, pints, quarts, or gallons. A calibration
factor and unit of measure are programmed during factory
test. Unlike mechanical registers, these units can be
electronically recalibrated in the field when necessary. A 6-
digit LC display, accurate to three decimal places, shows the
exact amount of fluid that has passed through the meter. The
entire register module is protected from normal wear and
tear by a rugged, shock resistant housing.
OPERATION
Industrial oval gear meters and nutating disc meters have
rotating magnets that cause the reed switches to send pulses
to the register when fluid passes the meter.
The register is in a sleep mode until it detects these pulses
caused by fluid going through the meter. The micro-
processor in the register then measures the flow and will
display the totalization and the flow rate of the fluid going
through the meter on the 6-digit display.
The registers batch totalizer is a 6-digit display with three
places of resolution after the decimal point.
If the total dispensed exceeds 999.999 then the display will
shift and only 2 digits will be displayed after the decimal
point, 9999.99 and will continue to shift to the maximum
value of 999999.
After reaching 999999 the batch totalizer will rollover to
0.000. The batch totalizer is reset to zero when the reset
button is depressed.
The register also has a resettable totalizer that requires that
the total and reset button both be depressed to reset (hold
the “Total” button, then press the “Reset” button to reset this
totalizer while resettable totalize is displayed). This would be
used for multiple batch totalization purposes.
The register´s life time totalizer is 11 digits and will either be
in gallons or liters based on the unit of measure selected.
Pushing and holding the total button while the life time
totalizer is displayed will display the full 11-digit life time
totalizer value.
ILR 750
ILR701T
ILR 701