Service manual

Bair Paws
®
Model 850 Warming Unit Service Manual
2
Important Information about the
Bair Paws Model 850 Warming Unit
Indications/Intended Use
e Model 850 warming unit is indicated for patient warming. is warming unit has been designed
for use with the Bair Paws warming gowns in all preoperative and postoperative settings.
Contraindication
Do not apply heat to lower extremities during aortic cross-clamping. ermal injury may occur if
heat is applied to ischemic limbs.
Warnings
• Do not warm patients with the warming unit’s hose alone. ermal injury may result.
Always connect the hose to a Bair Paws warming gown before providing patient
warming.
Do not use a forced-air warming device over transdermal medications; increased drug delivery, •
patient death, or injury may occur.
Use only Bair Paws warming gowns with the Model 850 warming unit. is warming unit has •
been designed to operate safely with Bair Paws warming gowns; use with other products may
cause thermal injury. To the full extent permitted by law, the manufacturer and/or importer
declines all responsibility for thermal injury resulting from the unit being used in conjunction
with products other than Bair Paws warming gowns.
e Model 850 warming unit is not intended for use in the operating room.•
Do not allow the warming unit’s hose to contact the patient’s skin during patient warming. •
ermal injury may result.
Position the temperature controller cord and the hose away from the patients neck or shoulders •
to avoid entanglement and/or injury.
Equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or •
with oxygen or nitrous oxide.