User's Guide

29R5911667 /03 AMM215WTTP
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally.
WARNING: Use of accessories, transducers and cables other than those specified or provided by
the manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
Electromagnetic emissions
The AMM215WTTP is intended for use in the electromagnetic environment specified below. The customer or
the user of the AMM215WTTP should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
Guidance
RF emissions
CISPR 11
Group 1 The AMM215WTTP uses RF
energy only for its internal function.
Therefore, its RF emissions are
very low and are not likely to cause
any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B The AMM215WTTP is suitable for
use in all establishments, including
domestic establishments and those
directly connected to the public low-
voltage power supply network that
supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class D
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Complies
This AMM215WTTP complies with appropriate medical EMC standards on emissions to, and interference
from surrounding equipment. Operation is subject to the following two conditions: (1) this device may not
cause harmful interference, and (2) this device must accept any interference received, including interference
that may cause undesired operation.
Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna or equipment.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The AMM215WTTP is intended for use in the electromagnetic environment specified below. The customer or
the user of the AMM215WTTP should assure that it is used in such an environment.
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic
environment guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast transient /
burst
IEC 61000-4-4
± 2 kV
100 kHz repetition frequency
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC 61000-4-5
± 0.5 kV, ± 1 kV, ± 2kV Mains power quality
should be that of a typical
Important Information