DICOM error threshold
5.16.1 About Embedded QA
About
Embedded QA allows you to run a display calibration or compliance test directly from the display using the
OSD menus described in the next sections. Embedded QA will use the front sensor / I-Guard to measure
the necessary luminance levels for either a calibration or compliance test. Various settings for both
actions
can be selected from the display’s OSD menu. The last results of both actions can be consulted from the
OSD.
Embedded QA or MediCal QAWeb?
Embedded QA is not a replacement for the Barco MediCal QAWeb solution.
Although Embedded QA is a reliable option to perform a simple calibration or compliance test, Barco still
highly recommends MediCal QAWeb as the solution of choice for calibration and QA. Medical QA Web
brings many benets such as centralized asset management, the ability to schedule tasks, remote
man-
agement, automated reporting, alerting and specic support of regional QA standards such as DIN 6868-
57, JESRA and AAPM TG18. That’s why MediCal QAWeb Agent acts as the master for all supported dis-
plays from the moment it is installed and running. MediCal QAWeb Agent will take over from Embedded
QA and overwrite any settings which were applied by Embedded QA.
5.16.2 DICOM status report
About DICOM status report
Following information is available:
Compliance status Shows if the current DICOM curve
is compliant or not.
Maximum error
Shows the maximum error of the
current D ICOM curve. This is the
deviation compared to a perfect
DICOM.
Error threshold
Shows the error threshold. This is
the maximum error allowed before
a DICOM calibration is required.
Time elapsed since latest
compliance check
Shows the backlight runtime since
last compliance check.
Display Function
Shows the current display
function.
Ambient light compensation
Shows the ambient light
compensation status.
Reading Room
Shows the selected reading room.
Luminance
Shows the measured luminance.
DICOM Compliance Status (status
since last compliance check)
Black luminance
Shows the measured backlight
luminance.
No calibration executed yet. No other information is visibleDICOM Calibration Status
Calibration executed. When the calibration is executed,
the following extra information
is shown: Backlight runtime
elapsed since latest calibration,
display function, ambient light
compensation, reading room.
26 K5902060 CORONIS 3MP LED DISPLAY 12/01/2015