Nio 3MP LED Display User Guide MDNC-3321 K5902067/03 26/11/2014
Barco nv President Kennedypark 35, 8500 Kortrijk, Belgium Phone: +32 56.23.32.11 Fax: +32 56.26.22.62 Support: www.barco.com/esupport Visit us at the web: www.barco.
Table of contents TABLE OF CONTENTS 1. Welcome! .......................................................................................... 3 1.1 1.2 1.3 About the product .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 3 Symbols . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 3 What’s in the box. .. .. .. ..
Table of contents 6.2 Repacking your display.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 35 7. Cleaning your display ........................................................................... 37 7.1 Cleaning instructions .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 37 8. Important information ......................................
1. Welcome! 1. WELCOME! 1.1 About the product Overview Thank you for choosing this Nio 3MP LED Display! Nio 3MP LED Display is an industry-standard 3MP color display system with LED backlights for dependable diagnostic viewing in high-bright color. Nio 3MP LED Display provides an effective display solution for a multitude of applications and modalities, including 3D PACS, 3D echo, ultrasound, orthopedic imaging, CAD, image fusion, nuclear medicine and PET.
1. Welcome! Note: gives extra information about the described subject Tip: gives extra advice about the described subject 1.3 What’s in the box Overview Your Nio 3MP LED Display comes with: • • • • • • • • • the display a system tilt and swivel foot this Nio 3MP LED Display user guide a system CD a DisplayPort cable a DVI cable an USB cable an AC power cord an external power supply If you ordered a Barco display controller, it’s also in the box together with its accessories.
2. Parts, controls and connectors 2. PARTS, CONTROLS AND CONNECTORS 2.1 Display front view Overview 1 2 3 4 5 6 Image 2-1 Front view 4 2 Left key Right key 5 Standby key Power LEDs 3 Menu key 6 Bottom downstream USB 1 The key icons are displayed above the keys, adapted to the function that it is used for (menu dependent). See "Navigating through the OSD menus", page 17.
2. Parts, controls and connectors 2.
2. Parts, controls and connectors Connector compartment cover open 1 2 3 4 5 6 7 Image 2-3 Rear view with open connector compartment cover. 1 +24 VDC power input 5 2 6 3 +5 VDC, 0.5A power output DisplayPort video input Ethernet connector (see note below) USB upstream connector 7 USB downstream connector 4 DVI video input The Ethernet connection is used for maintenance purposes and is not supported for user application.
2.
3. Display installation 3. DISPLAY INSTALLATION Prior to installing your Nio 3MP LED Display and connecting all necessary cables, make sure to have a suitable display controller physically installed in your computer. If you are using a Barco display controller, please consult the user guide delivered with it to do this. 3.1 Unlocking the height mechanism To remove the clip: 1. Position the display with its rear side facing you. 2. Pull the red clip out of the fixation hole in the foot. 3.
3. Display installation If, after installing the display of the system, you change the panel orientation while an image is on the screen, the result depends on the graphic board and the resolution of the image. In some cases the image will be rotated automatically, in other cases it will not be rotated (e.g., when pixels would be lost after rotation). If necessary, change the image resolution in the display control panel and restart the system after changing the orientation. 3.
3. Display installation 1. Connect the DVI or DP of the display controller to the DVI or DP connector of the display. 2. If you want to make use of the display’s USB upstream connector, connect a PC USB downstream connector to the display’s USB upstream connector. 3. If you have chosen to use the display’s USB downstream connector, connect a USB device to the downstream connector. 4. Connect the supplied external DC power supply to the +24 Vdc power input of the display. 5.
3. Display installation 3.5 Routing the cables & Reattach the connector compartment cover To route the cables 1. Route all connected cables through the cable routing channel in the stand of your display. Tip: The cable straps at the inside of the connector compartment allow you to fix the cables for better shielding of the cables. To Reattach the connector compartment cover 1. Reattach the connector compartment cover by sliding the cover’s bottom in position and then push the cover’s top.
3. Display installation WARNING: Use an arm that is approved by VESA (according to the VESA 100 mm standard). Use an arm that can support the weight of the display. Refer to the technical specifications of this display for the applicable weight. Overview The panel, standard attached to the tilt & swivel foot, is compatible with the VESA 100 mm standard. So it can be used with an arm stand according to the VESA 100 mm standard. Therefore, the tilt & swivel foot must be removed from the panel. 1.
3. Display installation 5. Remove the four screws fixing the foot while supporting the foot. 6. Attach the arm stand firmly to the panel using 4 screws M4 x 8 mm. 3.7 First time starting up Overview You are now ready to start up your Nio 3MP LED Display for the first time. 1. Switch on your Nio 3MP LED Display as described in "Standby switching", page 16. 2. Turn on the computer connected to your display.
4. Daily operation 4. DAILY OPERATION 4.1 Recommendations for daily operation Optimize the lifetime of your display Enabling the Display Power Management System (DPMS) of your display will optimize its diagnostic lifetime by automatically switching off the backlight when the display is not used for a specified period of time. By default, DPMS is enabled on your display, but it also needs to be activated on your workstation. To do this, go to “Power Options Properties” in the “Control Panel”.
4. Daily operation Maximize quality assurance The ’MediCal QAWeb’ system offers online service for high-grade Quality Assurance, providing maximum diagnostic confidence and uptime. Barco recommends to install MediCal QAWeb Agent and apply the default QAWeb policy at least. This policy includes calibration on regular intervals. Connecting to MediCal QAWeb Server offers even more possibilities. Learn more and sign up for the free MediCal QAWeb Essential level at www.barco.com/healthcare/qa. 4.
4. Daily operation 4.4 Bringing up the OSD menus How to bring up the OSD menus The OSD menu allows you to configure different settings to make your Nio 3MP LED Display fit your needs within your working environment. Also, you can retrieve general information about your display and its current configuration settings through the OSD menu. Bringing up the OSD menus can be done by: 1. If not already done so, switch on the display as previously described. 2. Illuminate the keys as previously described. 3.
4.
5. Advanced operation 5. ADVANCED OPERATION 5.1 OSD menu language About the OSD menu language By default, the OSD menu comes up in English. However, there’s a wide range of other languages available for the OSD menu of your Nio 3MP LED Display. To change the language of the OSD menu: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > User Interface > Menu menu. 3. Enter the Language submenu. 4. Select one of the available languages and confirm. 5.
5. Advanced operation 5.4 Key indicator lights About the key indicator lights By default, after lighting up, the key indicator lights will dim again if no further actions are taken within the following 5 seconds. However, this behavior can be changed so that the key indicator lights are always on or always off. To configure the key indicator lights 1. Bring up the OSD main menu. 2. Navigate to the Configuration > User Interface > Indicator Lights menu. 3. Enter the Keys submenu. 4.
5. Advanced operation 5.7 USB About USB By default, USB will be enabled. This will allow the connected PC to communicate directly over USB with the internal Nio 3MP LED Display processor. By disabling USB, communication between the internal Nio 3MP LED Display processor and the connected PC will not be possible. This does not disable the USB hub and still allows to make use of the display’s USB downstream connectors and any external devices connected to it (keyboard, mouse, ...).
5. Advanced operation 5.9 Hibernate About hibernate Enabling hibernation will not only switch off the backlight but will also force the display to disable other functionalities so that power consumption is further reduced to a minimum. This happens after a specific period of time which can be manually adjusted. Hibernate can only be enabled on your display when the DPMS mode is enabled first. Therefore, please refer to "DPMS mode", page 21 to do this. To enable/disable hibernation on your display: 1.
5. Advanced operation Diagnostic This mode provides the full calibrated luminance and is intended for using the display for diagnostic purposes. In this mode, the luminance is reduced to approximately half of the luminance. This is intended for using the display with office applications such as word processing. Text Please note that text mode is not persistent, once powered off, the unit will restart in diagnostic mode.
5. Advanced operation User Gamma 1.8 Gamma 2.2 This display function will be automatically selected when display functions are defined by MediCal QAWeb. Select one of these display functions in case the display is to replace a CRT display with a gamma of 1.8 or 2.2 respectively. To select a display function: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > Calibration menu. 3. Enter the Display Function submenu. 4. Select one of the available display functions and confirm. 5.
5. Advanced operation The available reading rooms for your Nio 3MP LED Display are: CR/DR/ MAMMO CT/MR/NM Staff Office Clinical Viewing Room Emergency Room Operating Room Corresponds to light conditions in diagnostic reading rooms for computed radiology, digital radiology or mammography. This setting has the lowest maximum ambient light. Corresponds to light conditions in diagnostic reading rooms for computed tomography, magnetic resonance or nuclear medicine scans.
5. Advanced operation 5.15.2 DICOM status report About DICOM status report Following information is available: DICOM Compliance Status (status Compliance status since last compliance check) Maximum error Reading Room Shows if the current DICOM curve is compliant or not. Shows the maximum error of the current DICOM curve. This is the deviation compared to a perfect DICOM. Shows the error threshold. This is the maximum error allowed before a DICOM calibration is required.
5. Advanced operation 5.15.4 DICOM calibration About DICOM calibration The DICOM calibration will add a correction to the current DICOM curve to approach the perfect DICOM curve as good as possible. To start DICOM calibration: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > Calibration > Embedded QA menu. 3. Select DICOM calibration to start the calibration. Warning:Pressing a key during calibration will abort the calibration, previous values will be restored.
5. Advanced operation To enable/disable image scaling: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > Image Source menu. 3. Enter the Scaling submenu. 4. Select Enabled/Disabled as desired and confirm. 5.17 Display orientation About Display orientation Your display automatically detects its physical orientation (portrait or landscape) and, by default, automatically adjusts the image orientation to this. This means that when your display is physically rotated, the image will rotate along.
5. Advanced operation however, features 10-bit encoding enabling it to better match the DICOM defined grayscale range. Such an extensive range is necessary to comply with the guidelines set forward by the latest medical guidelines. The available video encoding modes for your display are listed below. Please note that the display controller connected to your display might not always support all these video encoding modes.
5. Advanced operation To manually select a grayscale conversion mode: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > Image Source menu. 3. Enter the Grayscale Conversion submenu. 4. Select one of the available color conversion modes and confirm. 5.
5. Advanced operation To retrieve the status of your display: 1. Bring up the OSD main menu. 2. Navigate to the Status menu. 3. Enter the Display, Image Sources, Calibration or Connectivity submenu as desired.
5.
6. Repackaging instructions 6. REPACKAGING INSTRUCTIONS WARNING: Before repacking the display, follow the instruction to replace the protection buffer to prevent damage to the display.
6. Repackaging instructions 3 4 6.1 Protection buffer Display 7 Top box 8 Display controller box Replacing the protection buffer How to replace the protection buffer 1. Place the display on a stable surface. 2. Put the display in the lowest position and fasten the height mechanism, see "Unlocking the height mechanism", page 9 . 3. Very important: Tilt the panel away from the foot before changing the orientation. 4.
6. Repackaging instructions 6.2 Repacking your display How to repack your display 1. Place the empty bottom box on a stable surface. 2. Place the bottom buffer in the box. 3. Put the display together with the protection buffer (see "Replacing the protection buffer", page 34) in the dedicated cavity of the bottom buffer. 4. Put the top buffer on top of the display. 5. Slide the top box over the bottom box. 6. Use the three lock bayonets to lock the box.
6.
7. Cleaning your display 7. CLEANING YOUR DISPLAY 7.1 Cleaning instructions To clean the display Clean the display using a sponge, cleaning cloth or soft tissue, lightly moistened with a recognized cleaning product for medical equipment. Read and follow all label instructions on the cleaning product. In case of doubt about a certain cleaning product, use plain water.
7.
8. Important information 8. IMPORTANT INFORMATION 8.1 Safety information General recommendations Read the safety and operating instructions before operating the device. Retain safety and operating instructions for future reference. Adhere to all warnings on the device and in the operating instructions manual. Follow all instructions for operation and use. Electrical Shock or Fire Hazard To prevent electric shock or fire hazard, do not remove cover. No serviceable parts inside.
8. Important information To fully disengage the power to the device, please disconnect the power cord from the AC inlet. Power cords: • • • • • Utilize a UL-listed detachable power cord, 3-wire, type SJ or equivalent, 18 AWG min., rated 250 V min., provided with a hospital-grade type plug 5-15P configuration for 120V application, or 6-15P for 240V application. Do not overload wall outlets and extension cords as this may result in fire or electric shock. Mains lead protection (U.S.
8. Important information recycling of waste electrical and electronic equipment. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate these items from other types of waste and recycle them responsibly to promote the sustainable reuse of material resources. For more information about recycling of this product, please contact your local city office or your municipal waste disposal service. For details, please visit the Barco website at: http://www.
8. Important information 零件项目(名称) 有毒有害物质或元素 Component name Hazardous substances and elements 汞 六价铬 铅 镉 光盘说明书 Pb Hg Cd Cr6+ o o o o 多溴联苯 多溴二苯 醚 PBB PBDE o o CD manual O: 表示该有毒有害物质在该部件所有均质材料中的含量均在 SJ/T 11363-2006 标准规定的限量要求以下. O: Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement in SJ/T11363-2006. X: 表示该有毒有害物质至少在该部件的某一均质材料中的含量超出 SJ/T 11363-2006 标准规定的 限量要求.
8. Important information 8.3 Regulatory compliance information Indications for use The Nio 3MP LED Display is intended to be used as a tool in displaying and viewing digital images (excluding digital mammography) for review and analysis by trained medical practitioners. Caution (USA): Federal law restricts this device to sale by or on the order of a physician. (Details & exemptions are in the Code of Federal Regulations Title 21, 801 Part D).
8. Important information Emissions test Compliance RF emissions Class B CISPR 11 Harmonic emissions Class D IEC 61000-3-2 Voltage fluctuations/ flicker emissions Electromagnetic environment – Guidance The Nio 3MP LED Display is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
8. Important information Immunity test Electromagnetic environment – Test levels guidance Voltage dips, short < 5% U T 1(> 95% dip in < 5% U T (> 95% dip in Mains power quality interruptions and voltage U T) for 0.5 cycle should by that of a U T) for 0.5 cycle variations on power typical commercial or 40% U T (60% dip in U T) 40% U T (60% dip in U T) supply input lines hospital environment.
8. Important information Immunity test IEC 60601 Compliance level Test levels Electromagnetic environment – guidance survey,3 should be less than the compliance level in each frequency range.4 Interference may occur in the vicinity of equipment marked with symbol: At 80 MHz and 800 MHz, the higher frequency range applies. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
8. Important information These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection form structures, object and people. 8.
8. Important information Indicates the DisplayPort connectors on the device Indicates the manufacturing date Indicates the temperature limitations6 for the device to safely operate within specs Indicates the device serial no.
8. Important information Indicates the maximum number of boxes to be stacked on each other. 15 n Indicates that the box should be carried with two persons. 20 - 30 Kg Indicates that the box should not be cut with a knife, a cutter or any other sharp object. yy °C +60 xx °C -20 yy % 85 5x % yyy kPa 106 Indicates the temperature limits to which the medical device can be safely exposed when being stored.
8. Important information 8.7 Technical specifications Overview Product acronym MDNC-3321 Screen technology UA-SFT (ultra advanced - super fine technology)+H106 Active screen size (diagonal) 540 mm (21.3”) Active screen size (H x V) 324,9 mm x 433,2 mm (12.79” x 17.1”) 4:3 Aspect ratio (H:V) Resolution Pixel pitch 3MP (2048 x 1536) 0.
8.
8. Important information BE LIABLE FOR DIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGES. MORE INFORMATION/DETAILS IS TO BE FOUND IN EACH SPECIFIC OPEN SOURCE LICENSE. Copyright on each Open Source Software component belongs to the respective initial copyright holder, each additional contributor and/or their respective assignee(s), as may be identified in the respective documentation, source code, README file, or otherwise.