Nio User Guide Supported displays: E-3620, MDNG-5121, MDNG-2121, MDNC-3121, MDNG-6121, MDNC-2121
(This page intentionally left blank.
Welcome! Nio is Barco's flexible, industry-standard display solution for diagnostic imaging. Presenting an exceptional combination of innovation and performance, Barco's next-generation Nio display system improves your diagnostic reading routine for a multitude of medical imaging applications.
(This page intentionally left blank.
Table of contents What's in the box.............................................................7 Parts, controls and connectors ........................................8 Front view ............................................................................................... 8 Rear view ................................................................................................ 9 Display installation ........................................................12 Unlocking the height mechanism ...........
Concepts......................................................................................... 35 Display function selection ............................................................ 36 ALC & DICOM Options.................................................................... 37 Switching DPMS on/off........................................................................ 40 Cleaning your display ....................................................41 Front glass ..........................................
What's in the box Your BARCO NIO Display comes with: • this BARCO NIO Display guide • a system CD • a DVI cable • a DisplayPort cable (MDNC-2121 only) • a USB cable • a set of AC power cords • an external power supply If you ordered a Barco display controller, it’s also in the box together with its accessories. A dedicated user guide is available on the system CD. Keep your original packaging. It is designed for this display and is the ideal protection during transport.
Parts, controls and connectors Front view 1 BAR CO 3 2 Figure 1: Front side 8 1 Power LED 2 3 Control wheel Nio USB downstream port
Rear view REAR VIEW OF E-3620, MDNG-5121 1 2 4 5 6 7 8 9 3 Figure 2: Rear side 1 Connector compartment cover 2 Tilt & swivel foot cover 3 Tilt & swivel foot 4 Slot for security cable 5 DVI (digital) video input 6 USB upstream port 7 USB downstream port 8 DC power input 9 Tilt & swivel foot clip Nio 9
REAR VIEW OF MDNC-3121, MDNG-2121 AND MDNG-6121 1 2 4 56 7 8 9 10 3 Figure 3: Rear side 10 1 Connector compartment cover 2 Tilt & swivel foot cover 3 Tilt & swivel foot 4 DC power input 5 +5 Vdc 6 Slot for security cable 7 DVI (digital) video input 8 USB upstream port 9 USB downstream port 10 Tilt & swivel foot clip Nio
REAR VIEW OF MDNC-2121 1 2 4 5 6 7 8 9 10 3 Figure 4: Rear side 1 Connector compartment cover 2 Tilt & swivel foot cover 3 Tilt & swivel foot 4 Slot for security cable 5 DisplayPort video input 6 DVI (digital) video input 7 USB upstream port 8 USB downstream port 9 DC power input 10 Tilt & swivel foot clip Nio 11
Display installation Prior to installing your BARCO NIO Display and connecting all necessary cables, make sure to have a suitable display controller physically installed in your computer. If you are using a Barco display controller, please consult the dedicated user guide available on the system CD. Unlocking the height mechanism In the factory, the height-positioning system in the display foot is blocked with a red clip to prevent damage during transportation.
Adjusting the display position You can change the orientation of the panel at any time, but it is more convenient to select landscape or portrait orientation before connecting the cables..
TO CHANGE THE PANEL ORIENTATION 1 Stand at the front side of the panel and take the panel at both sides. 2 Very important: Til the panel before changing the orientation. Should you change the panel orientation without tilting it first, you might irreversibly damage the tilt & swivel mechanism.
3 To change from portrait to landscape, turn the panel counterclockwise. BAR CO Figure 9: To rotate the panel from portrait to landscape 4 To change from landscape to portrait, turn clockwise.. If, after installing the display or the system, you change the panel orientation while an image is on the screen, the result depends on the graphic board and the resolution of the image. In some cases the image will be rotated automatically, in other cases it will not be rotated (e.g.
Connecting the signal cables TO CONNECT THE SIGNAL CABLES TO THE DISPLAY: To get access to the connectors, remove the connector compartment cover by pulling down the 2 clips at the top of the cover. The location of the connectors depends on the display type: 2 1 3 4 Figure 10: Inputs E-3620, MDNG-5121 4 2 3 1 Figure 11: Inputs MDNC-3121, MDNG-2121 and MDNG-6121 1 2 3 4 Figure 12: Inputs MDNC-2121 • Connect one end of the DVI/DisplayPort cable to the DVI/ DisplayPort input of the display (1).
Routing the cables • Bind the cables in the connector compartment together with the cable tie inside the connector compartment. • Put the connector compartment cover back on the display. Pay attention that the signal cables are positioned under the bulge in the cover. • Push the cables into the clips on the rear of the tilt & swivel foot. • Bind the cables together above and under the foot, by means of the 2 velcro strips attached to the inside of the foot cover (packed inside the accessory box).
3 Slide the cover upward while moving the lower side of the cover towards the foot. 4 Press the cover to the foot so that it makes a clicking sound. VESA-mount installation The panel, standard attached to the tilt & swivel foot, is compatible with the VESA 100 mm standard. So it can be used with an arm stand according to the VESA 100 mm standard. Therefore, the tilt & swivel foot must be removed from the panel.. • Use an arm that is approved by VESA (according to the VESA 100 mm standard).
3 Remove the small screw (A) fixing the small plastic cover on top of the foot. Next, remove the small cover itself. B A Figure 14: Display with tilt & swivel foot cover removed 4 Unscrew the 2 screws fixing the round plastic cover (B). 5 Lift up the round plastic cover. 6 Remove the four screws fixing the foot while supporting the foot. 7 Attach the arm stand firmly to the panel using 4 screws M4 x 8 mm.
Daily operation Recommendations for daily operation OPTIMIZE THE LIFETIME OF YOUR DISPLAY Enabling the Display Power Management System (DPMS) of your display will optimize its diagnostic lifetime by automatically switching off the backlight when the display is not used for a specified period of time. By default, DPMS is enabled on your display, but it also needs to be activated on your workstation. To do this, go to “Power Options Properties” in the “Control Panel”.
performance of the product. To ensure optimal product quality, Barco applies strict selection criteria for its LCD panels. To learn more about LCD technology and missing pixels, consult the dedicated white papers available at www.barco.com/ healthcare. ENHANCE USER COMFORT Every Barco multi-head display system is color matched with the highest specification in the market. Barco recommends keeping color-matched displays together.
Stand-by switching The connected power supply also provides a switch that can be used to turn the power completely off. To use the display, please make sure to switch on this power supply. This can be done by pushing the on/off switch on the power supply into the "|" position. ON E-3620 AND MDNG-5121 DISPLAYS: When the display is on and no on-screen display is visible, push and hold the control wheel at the front for a few seconds to switch the display in stand-by. The LED turns orange.
ON MDNC-2121, MDNC-3121, MDNG-2121 AND MDNG6121 DISPLAYS: 1 When the display is on and no on-screen display is visible, push the control wheel at the front shortly. 2 An on-screen message appears, asking to press to confirm. 3 Push the control wheel once more to confirm and switch to standby. 4 The LED turns orange. BAR CO Push twice Figure 17: Location of the control wheel 5 When the display is in stand-by, press the control wheel shortly to switch it back on.
TO DISABLE USER CONTROLS ON E-3620 AND MDNG-5121: 1 Rotate the control wheel to display the on-screen display. The Main Menu appears. 2 Rotate the control wheel to select Settings. 3 Press the control wheel to enter the Settings menu. SETTINGS DPMS Power LED User Controls EXIT On On On Figure 18: Settings menu (example) 4 Rotate the control wheel to select User Controls. 5 Press the control wheel to switch from “On” to “Off”. 6 Exit the menus.
7 Press the control wheel to confirm and return to the Settings menu. 8 Exit the menus. TO ENABLE USER CONTROLS: 1 Make sure the on-screen display is not on the screen. 2 Do not use the control wheel for at least 3 seconds. 3 Only for MDNC-2121, MDNC-3121, MDNG-2121 and MDNG-6121: Press and hold the control wheel for at least 5 seconds. 4 Rotate the control wheel 1 step clockwise. 5 Press the control wheel 2 times. 6 Rotate the control wheel 1 step counterclockwise. .
Navigating through the OSD menus 1 Rotate the control wheel to move up and down through the menus. 2 To go into a submenu, rotate the control wheel to highlight the submenu. Next, press the control wheel to go into the submenu. 3 When you press on Exit, the OSD exits the submenu and jumps to the previous (higher-level) menu. When you press on Exit in the main menu, the OSD is hidden.
Changing values in the OSD menus 1 Navigate through the menus and highlight the parameter you wish to adjust. 2 Press the control wheel to go into edit mode. 3 Rotate the wheel to change the value. 4 Press the wheel to enter the new value and return to the menu.
Advanced operation On-screen display (OSD) Input Selection menu* Auto Detects the incoming video signal automatically and displays it on the screen. DVI Always displays the DVI signal on the sreen. When no DVI signal available, there will be no image on the screen. DP Always displays the DisplayPort signal on the screen. When no DP signal available, there will be no image on the screen.
Luminance and color menu In grayscale Nio displays this menu is called “Luminance”. Measured luminance** Indicates the actual luminance measured by the internal sensor. This is a read-only value. It is expressed in a percentage of the calibrated value (100% is calibrated). Luminance target Allows to manually adjust the luminance target. See note below. Color target* Allows to select from a list of factorydefined and user-defined color targets.
When you change the luminance target, the display will adjust its backlight to reach the target. This can be seen in the Measured luminance line. When the luminance target cannot be reached, e.g., due to aging of the backlight, the Measured luminance line changes to Minimum value reached or Maximum value reached. Display Function menu* Display function Allows to select from a list of pre-defined display functions. If the DICOM DF is selected, an additional function is available.
ALC & DICOM OPTIONS SUBMENU* Correction value Shows the ambient light correction value that is taken into account in the calculation of the display function, measured in lux. DICOM offset The DICOM DF is recalculated at the moment the DICOM Offset is changed, taking the new DICOM Offset into account. • When set to “Dark Room”, the ambient light is not taken into account. • When set to “Preset”, a preset ambient light value determined by the selected reading room is taken into account.
READING ROOM DEFINITION SUBMENU* Reading room Indicates the reading room type you are editing. You can select another room to edit by rotating the control wheel. Max. ambient light This indicates the maximum ambient light that corresponds to the selected reading room. If the measured ambient light is higher than the value entered here, you should take measures to darken the room or select another reading room type. Preset corr.
Settings menu DPMS Allows to switch the display power management system on/off. See note below. Power LED Allows to switch the power LED’s on state on/off. The LED's orange DPMS state is not influenced by this setting. So, when the display goes into power-saving mode, the LED will turn orange, even if it was switched off by this setting User controls Allows to disable the control wheel on the front. When switched off, the user cannot display the OSD unit the user controls are enabled again.
*Only in MDNC-2121, MDNC-3121, MDNG-2121 and MDNG-6121. The DPMS system will power down the display when the connected computer is left idle for a certain time. Barco recommends to switch DPMS on and to use a good Windows screen saver to prevent image burn-in (image retention) on the LCD panel.
Changing Display Functions Concepts Display function selection is possible only in MDNC-3121, MDNG-2121 and MDNG-6121 displays. • The Display Function menu lets you select a display function (DF). • When you select the DICOM DF you can define the DICOM DF to be recalculated taking the ambient light into account. • In this case, the recalculation can be based on an illuminance of 0 lux (“Dark Room”) or a preset value (“Preset”) determined by the selected reading room.
Display function selection TO SELECT A DISPLAY FUNCTION (DF): 1 When there is no OSD on the screen, rotate the control wheel to display the OSD. The main menu appears. 2 Rotate the control wheel to select the Display Function menu. Press the control wheel shortly to enter the menu. 3 Rotate the control wheel to select Display Function. Press the control wheel shortly to go into edit mode. 4 Rotate the control wheel to select the desired DF. 5 Press the control wheel shortly to activate the new selection.
User This is a user-definable DF. Use MediCal QAWeb to define the DF. ALC & DICOM Options DICOM offset INTRODUCTION The DICOM DF will be recalculated only when another DICOM Offset is selected. The DICOM DF will be recalculated based on the new offset setting. TO SELECT A DICOM OFFSET: 1 Select the DICOM display function. See To select a display function (DF):, on page 36. 2 Enter the ALC & DICOM submenu. 3 In the ALC & DICOM Options submenu, rotate the control wheel to select DICOM Offset.
YOU CAN SELECT THE FOLLOWING OFFSETS: Dark room The DICOM DF will be recalculated taking an ambient light value of 0 lux into account. This means that the measured ambient light does not influence the DICOM DF. Preset The DICOM DF will be recalculated taking a preset ambient light value into account. This value is determined by the selected Reading Room and can be adjusted manually within certain limits. Reading room selection This function is available only when the DICOM display function is selected.
CT/MR/NM Corresponds to light conditions in diagnostic reading rooms for computed tomography, magnetic resonance or nuclear medicine scans. Office Corresponds to light conditions in office rooms. Clinical Corresponds to light conditions in diagnostic reading rooms for clinical viewing. Emergency Corresponds to light conditions in emergency rooms. Operating Corresponds to light conditions in operating rooms. This setting has the highest maximum ambient light.
5 To modify the maximum ambient light for this reading room, rotate the control wheel to select Max. Ambient Light. Press the control wheel shortly to go into edit mode. Rotate the control wheel to change the value. Press the control wheel shortly to save the changes. You can change this value within certain limits determined by the selected reading room. 6 To modify the Preset correction value for this reading room, rotat e the control wheel to select Preset Corr. Value.
Cleaning your display Front glass Clean the front glass using a sponge, cleaning cloth or soft tissue, lightly moistened with one of the following tested products: • Flux • Pril • Windex Glass Plus • Ajax glass cleaner • Bohle glass cleaner • Sidolin glass cleaner • Mr. Proper • 25% ethanol solution (EtOH) In case none of the above cleaning products is available, use plain water. Take care not to damage or scratch the front glass.
Do not use following products: • Alcohol/solvents at higher concentration > 5% • Strong alkalis lye, strong solvents • Acid • Detergents with fluoride • Detergents with ammonia • Detergents with abrasives • Steel wool • Sponge with abrasives • Steel blades • Cloth with steel thread Cabinet 1 Clean the cabinet using a soft cotton cloth, lightly moistened with a recognized cleaning product for medical equipment.
Important information Safety information GENERAL RECOMMENDATIONS Read the safety and operating instructions before operating the equipment. Retain safety and operating instructions for future reference. Adhere to all warnings on the equipment and in the operating instructions manual. Follow all instructions for operation and use.
POWER CONNECTION - EQUIPMENT WITH EXTERNAL 12 VDC POWER SUPPLY • Power requirements: The equipment must be powered using the delivered medical approved 12 VDC power supply. • The medical approved DC power supply must be powered by the AC mains voltage (protective earth terminal). POWER CONNECTION - EQUIPMENT WITH EXTERNAL 24 VDC POWER SUPPLY • Power requirements: The equipment must be powered using the delivered medical approved 24 VDC SELV power supply.
WATER AND MOISTURE Never expose the equipment to rain or moisture. Never use the equipment near water - e.g. near a bathtub, washbasin, swimming pool, kitchen sink, laundry tub or in a wet basement. VENTILATION Do not cover or block the ventilation openings in the cover of the set. When installing the equipment in a cupboard or another closed location, heed the necessary space between the set and the sides of the cupboard.
Environmental information DISPOSAL INFORMATION (WASTE ELECTRICAL AND ELECTRONIC EQUIPMENT) This symbol on the product indicates that, under European Directive 2002/96/EC governing waste from electrical and electronic equipment, this product must not be disposed of with other municipal waste. Please dispose of your waste equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment.
中国大陆 ROHS (CHINESE MAINLAND ROHS) 根据中国大陆 《电子信息产品污染控制管理办法》(也称为中国大 陆 RoHS),以下部分列出了 Barco 产品中可能包含的有毒和 / 或有 害物质的名称和含量。中国大陆 RoHS 指令包含在中国信息产业部 MCV 标准:“ 电子信息产品中有毒物质的限量要求 " 中。 According to the "China Administration on Control of Pollution Caused by Electronic Information Products" (Also called RoHS of Chinese Mainland), the table below lists the names and contents of toxic and/or hazardous substances that Barco's product may contain.
电源供应器 Power supply unit 文件说明书 Paper manuals 光盘说明书 CD manual X O O O O O O O O O O O O O O O O O O: 表示该有毒有害物质在该部件所有均质材料中的含量均在 SJ/T 11363-2006 标准规定的限 量要求以下 . O: Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement in SJ/T11363-2006. X: 表示该有毒有害物质至少在该部件的某一均质材料中的含量超出 SJ/T 11363-2006 标准规 定的限量要求 .
Regulatory compliance information INDICATIONS FOR USE: E-3620, MDNG-2121, MDNC-3121, MDNC-2121 The products are intended to be used in displaying and viewing digital images, for review and analysis by trained medical practitioners. These devices must not be used in primary image diagnosis in mammography. Caution (USA): Federal law restricts this device to sale by or no the order of a physician or a practitioner trained on its use.
• Reorient or relocate the receiving antenna. • Increase the separation between the equipment and receiver. • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. • Consult the dealer or an experienced radio/TV technician for help. CANADIAN NOTICE This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme à la norme NMB-001 du Canada.
Explanation of symbols On the display or power supply, you may find the following symbols (nonrestrictive list): 0120 Indicates compliance to the essential requirements of the Directive 93/42/EEC Indicates compliance with Part 15 of the FCC rules (Class A or Class B) Indicates the display is approved according to the UL regulations or or Indicates the display is approved according to the c-UL regulations Indicates the display is approved according to the DEMKO regulations S&E Indicates the display is
Consult the operating instructions Indicates this apparatus must not be thrown in the trash but must be recycled, according to the European WEEE (Waste Electrical and Electronic Equipment) directive Legal disclaimer DISCLAIMER NOTICE Although every attempt has been made to achieve technical accuracy in this document, we assume no responsibility for errors that may be found. Our goal is to provide you with the most accurate and usable documentation possible; if you discover errors, please let us know.
Technical specifications NIO 2MP Product acronym MDNG-2121 Screen technology TFT AM LCD Dual Domain IPS Active screen size 540 mm (21.3") (diagonal) Active screen size (H x V) 432 x 324 mm (17.0 x 12.8") Aspect ratio (H:V) 4:3 Resolution 2MP (1600 x 1200) Pixel pitch 0.
Response time (Tr + Tf) 23 ms Scanning frequency (H; V) 15-129 kHz; 24-100 Hz Housing color Gray Video input signals DVI-D Dual Link USB ports 1 upstream (endpoint), 2 downstream USB standard 2.
Swivel -45° / +45° Pivot 90° Mounting standard VESA (100 mm) Screen protection Protective, non-reflective PMMA cover Recommended modalities CT, MR, US, DR, CR, NM, Film Certifications BSMI, CCC - GB4943-2001 - GB9254-1998 - GB92542008 - GB17625.1-2003, DEMKO EN60601-1:1990 + A1:1993 + A2:1995, IEC 60601-1:1998 + A1:91 + A2:95, UL 60601-1 1st edition, CAN/CSA-CS22.2 No.601.
NIO 3MP Product acronym E-3620 (MA) Screen technology TFT AM LCD Dual Domain IPS Active screen size 528 mm (20.8") (diagonal) Active screen size (H x V) 424 x 318 mm (16.7 x 12.5") Aspect ratio (H:V) 4:3 Resolution 3MP (2048 x 1536) Pixel pitch 0.
Housing color Gray Video input signals DVI-D Dual Link USB ports 1 upstream, 2 downstream USB standard 1.
Mounting standard VESA (100 mm) Screen protection Protective, non-reflective PMMA cover Recommended modalities CT, MR, US, DR, CR, NM, Film Certifications CCC - GB4943-2001 - GB9254-1998 (Class A) GB9254-2008 (Class A) - GB17625.1-2003, DEMKO EN60601-1:1990 + A1:1993 + A2:1995 + A11: 1993 + A12: 1993 + A13:1996, IEC 60601-1:1988 + A1:1991 + A2:1995 2nd edition, UL 60601-1 1st edition 2006-0426, CAN/CSA-C22.2 No.601.
NIO 5MP Product acronym MDNG-5121 Screen technology TFT AM LCD Dual Domain IPS Active screen size 541 mm (21.3") (diagonal) Active screen size (H x V) 422 x 338 mm (16.5 x 13.3") Aspect ratio (H:V) 5:4 Resolution 5MP (2560 x 2048) Pixel pitch 0.
Housing color Gray Video input signals DVI-D Dual Link USB ports 1 upstream (endpoint), 2 downstream USB standard 1.
Mounting standard VESA (100 mm) Screen protection Protective, non-reflective PMMA cover Recommended modalities CT, MR, US, DR, CR, NM, Film Certifications BSMI, CCC - GB4943-2001 - GB9254-1998 (Class A) GB9254-2008 (Class A) - GB17625.1-2003, DEMKO EN60601-1:1990 + A1:1993 + A2:1995, IEC 606011:1998 + A1:1991 + A2:1995 2nd edition, IEC 60601-1 (ed.2) am1 & am2, UL 60601-1 1st edition 2006-0426, CAN/CSA-C22.2 No.601.
NIO COLOR 3MP HB Product acronym MDNC-3121 Screen technology TFT AM Color LCD IPS Active screen size 540 mm (21.3") (diagonal) Active screen size (H x V) 433 x 325 mm (17.0 x 12.8") Aspect ratio (H:V) 4:3 Resolution 3MP (2048 x 1536) Pixel pitch 0.
Video input signals DVI-D Dual Link USB ports 1 upstream (endpoint), 2 downstream USB standard 2.
Screen protection Protective, non-reflective PMMA cover Recommended modalities CT, MR, US, DR, CR, NM, Film Certifications BSMI, CCC - GB4943-2001 - GB9254-2008 GB17625.1-2003, DEMKO EN60601-1:1990 + A1:1993 + A2:1995 + A13:1996, IEC 60601-1:1988 + A1:1991 + A2:1995 2nd edition, UL 60601-1 1st edition 2006-0426, CAN/CSA-C22.2 No.601.1-M90 2005, FDA510K, RoHs, MDD, WEEE, CE 0120, VCCI, FCC Class B, KETI, IEC 60601-1:2005, EN 60601-1-2:2006, ANSI/AAMI ES 60601-1:2005, CAN/CSA-C22.2 No.
NIO 5MP Product acronym MDNG-6121 Screen technology TFT AM LCD Dual Domain IPS Active screen size 541 mm (21.3") (diagonal) Active screen size (H x V) 432,6 x 324,5 mm (17.0 x 12.8") Aspect ratio (H:V) 4:3 Resolution 5MP (2800 x 2096) Pixel pitch 0.
Housing color Gray Video input signals DVI-D Dual Link USB ports 1 upstream (endpoint), 2 downstream USB standard 2.
Mounting standard VESA (100 mm) Screen protection Protective, non-reflective PMMA cover Recommended modalities CT, MR, US, DR, CR, NM, Film Certifications BSMI, CCC - GB4943-2001 - GB9254-1998 - GB92542008 - GB17625.1-2003, DEMKO EN60601-1:1990 + A1:1993 + A2:1995 + A13:1996, IEC 60601-1:1988 + A1:1991 + A2:1995 2nd edition, UL 60601-1 1st edition 2006-04-26, CAN/CSA-C22.2 No.601.
NIO COLOR 2MP HB Product acronym MDNC-2121 Screen technology TFT AM Color LCD IPS Active screen size 541 mm (21.3") (diagonal) Active screen size (H x V) 432 x 324 mm (17.0 x 12.8") Aspect ratio (H:V) 4:3 Resolution 2MP (1600 x 1200) Pixel pitch 0.2700 mm (0.
Video input signals DVI-D Single Link (1x), DisplayPort (1x) USB ports 1 upstream (endpoint), 2 downstream USB standard 2.
Screen protection Protective, non-reflective PMMA cover Recommended modalities CT, MR, US, DR, CR, NM, Film Certifications UL60601-1, CB60601-1, CSA C22.2 No 601.1, EN606011-2, EN55011 Limit B, ICES-001 B, EN61000-3-2/3, EN61000-4-2/3/4/5/6/8/11, VCCI, CE0120, KETI, CCC, BSMI, Demko, FCCB, FDA 510K, CMDCAS, MDD, WEEE, RoHs, IEC 60601-1:2005, EN 60601-1-2:2006, ANSI/AAMI ES 60601-1:2005, CAN/CSA-C22.2 No.
Nio 71