Concepts
The Display Function menu lets you select a display function (DF).
When you select the DICOM DF you can define the DICOM DF to be
recalculated taking the ambient light into account.
In this case, the recalculation can be based on an illuminance of 0
lux (“Dark Room”) or a preset value (“Preset”) determined by the
selected reading room.
In the latter case you must select a reading room that corresponds
to the type of room the display is installed in. E.g. if the display is
installed in a room where CT scans are observed, the “CT/MR/NM”
reading room is the preferred selection.
The reading room is defined by 2 parameters:
Maximum Ambient Light defines the maximum light
allowed in this type of room. This value can be adjusted
within certain limits determined by the selected reading
room.
Preset Correction Value is the predefined correction value
for this reading room. This value can be adjusted within
certain limits determined by the selected reading room.
The reading room parameters are pre-defined in the display
according to the AAPM (American Association of Physicists in
Medicine). However they can be changed within certain limits.
Display function selection is possible only in MDNC-3121,
MDNG-2121 and MDNG-6121 displays.