User guide
38 Nio
YOU CAN SELECT THE FOLLOWING OFFSETS:
Reading room selection
This function is available only when the DICOM display function is
selected.
TO SELECT A READING ROOM SETTING:
1 Select the DICOM display function. See To select a display
function (DF):, on page 36.
2 Enter the ALC & DICOM Options submenu.
3 In the ALC & DICOM Options submenu, rotate the control wheel to
select Reading Room. Press the control wheel shortly to go into
edit mode.
4 Rotate the control wheel to select the reading room that
corresponds to the room type the display is installed in.Return to
the menu.
5 You can select the following reading room types:
Dark room The DICOM DF will be recalculated taking
an ambient light value of 0 lux into
account.
This means that the measured ambient
light does not influence the DICOM DF.
Preset The DICOM DF will be recalculated taking a
preset ambient light value into account.
This value is determined by the selected
Reading Room and can be adjusted
manually within certain limits.
CR/DR/MAMMO Corresponds to light conditions in
diagnostic reading rooms for computed
radiology, digital radiology or
mammography. This setting has the lowest
maximum ambient light.