Specifications

Air Purge
1. If the optional remote dispenser is connected to
the EASYpure II UV/UF, place the draw-off trig-
ger into the locked position by depressing the
button on the right side of the dispenser at the
same time as the trigger is depressed.
OR
If there is no remote dispenser being utilized,
ensure that the hose barb is attached to the
draw-off valve fitting and depress the draw-off
valve lever to open the draw-off valve.
2. Place a suitable container under the opened
valve to collect water while purging air.
3. Connect the water supply to the EASYpure II
UV/UF and open the customer supplied shut-off
valve.
4. Turn the power switch on.
5. Press the START/STOP switch on the control
panel.
6. When air is purged and a steady stream of
water is flowing from the opened valve, close
the valve.
7. Press the FLUSH switch on the control panel
twice to begin an extended flush to drain.
8. When the extended flush is completed, the sys-
tem will return to the Run Mode. Press the
START/STOP switch to place into the Idle Mode.
9. Refer to the Sanitization section and sanitize
the EASYpure II UV/UF.
21
I
NSTALLATION
Caution
Do not allow EASYpure II UV/UF to
operate unless water is available to
unit.
Note
See the Controls and Normal
Operation section to familiarize your-
self with the controls of the unit.