Specifications
Sanitization
Your EASYpure II UV/UF has been shipped with a sanitiza-
tion syringe (Cat. No. CMX25). When the system is new or
whenever new cartridges or a new ultrafilter are installed, the
system should be sanitized. Sanitize your EASYpure II
UV/UF as follows:
1. If the 0.2 micron final filter is installed, remove it
and attach the hose barb that came installed on the
unit.
2. Turn the system power off.
3. Disconnect the feed water supply at the inlet quick
disconnect connection.
4. Ensure that the remote dispenser (if applicable) is
open and pointed to drain. If no remote dispenser
is present, ensure that the unit draw-off valve is
open and a container is placed under the draw-off
valve.
5. Remove the sanitization syringe from wrapper.
6. Remove the luer cap on the sanitization block,
located on the inside left of the cartridge area. (See
Fig. 13)
7. Remove the syringe cap and attach syringe to the
luer fitting on the sanitization block.
8. Slowly inject the solution into the system. Some
water will be displaced from the remote dispenser
or unit draw-off valve. Avoid injecting air into the
system.
9. Remove the syringe and replace the luer cap on
the sanitization block.
10. Close the remote dispenser or unit draw-off valve.
11. Reconnect the feed water supply removed in step
3.
12. Allow the system to stand for at least 1 hour.
13. If using an optional remote dispenser, turn system
power on and from “IdL”, start the system by press-
ing the START switch and allow to run for 3-5 sec-
onds. Press the STOP switch and leave the sys-
tem in “IdL” for 1 hour. If no remote dispenser is
used, proceed to step 14.
Maintenance and Servicing
Feedwater Inlet
Figure 13: Sanitization
26
Warning
Disconnect from the power supply prior
to maintenance and servicing.
Do not disassemble water lines or re-
move cartridges where spilled water
could contact equipment that requires
electrical service. Electrical shock haz-
ard could result.
Refer servicing to qualified personnel.