Instructions

This product is used to measure the height size and scribe.
Clean the bottom of base and the scriber, install the scriber, put the height gage on an
inspection plate.
The height gage should be set zero before measuring:
--- Make sure the bottom of the base and the inspection plate is clean, without burrs and
scratches and other effects.
--- Lower the scriber, make the measuring surface and the bottom of the base at the same
level, check whether the reading is zero. lf there is deviation, move the beam to set zero.
--- How to set zero:
Loosen locking nut, rotate adjustment nut, and can move beam up and down, adjust zero
position, so that the zero position of vernier coincides with the beam.
--- Tighten the locking nut to finish the calibration.
During reading, the sight is perpendicular to the scale to avoid parallax reading. The reading
is the sum of beam and vernier. The reading results are as follows:
www.accud.com
9
8
7
6
10
2
1
5
4
3
Dial test indicator holder, for holding dial test indicator.
Applicable products
186-012-11 / 186-024-11 / 186-040-11
Graduation: 0.02mm/0.001''
Accuracy: ±0.04mm(Range: 0-300mm)
±0.05mm(Range: 300-600mm)
±0.07mm(Range: 600-1000mm)
Base
Scriber
Clamp
Vernier
Beam
1
2
3
4
5
Magnifier
Locking screw
Fine adjustment frame
Locking nut
Adjustment nut
6
7
8
9
10
VERNIER HEIGHT GAGE
Check the bottom of the base and the inspection plate is clean, without burrs and scratches
and other effects.
When measuring the height, raise the measuring surface of the scriber slightly higher than
the measured height of the workpiece, and then use the fine adjustment frame to make the
measuring surface of the scriber contact with the measured surface of the workpiece and
then get the reading. Be careful of the measuring force to avoid the measurement error.
1. Precautions before use
2. Measurement
3. Reading
4. Optional accessory
5. Notice
Beam reading: 20mm
Vernier scale reading: 0.30mm
Reading: 20.30mm
With series 261/267
With series 267/222
MNE-186-11
Use dial test indicator (precision dial test indicator or dial indicator) instead of scriber to
position to improve measurement accuracy. Install dial test indicator into the holder. Firstly
make the dial test indicator touch with the inspection plate and set zero, get the reading a1
from the height gage, and then move the beam to make the contact point contact with the
upper surface of the workpiece and set zero, get the reading a2, the height of the workpiece
The scriber is made of carbide and can be used for scribing. When scribing, use the fine
adjustment frame to accurately align the required size with the height of the scribe line. then
tighten the locking screw and scribe. When scribing, the base should fit the inspection plate
well and can move smoothly. The contact pressure between the scriber and the surface of the
workpiece should be appropriate to ensure that the scribing is clear but without deep marks
When move the gage, you should hold the base with one hand and hold the beam with the
other hand.
After use, clean and put the gage back to the packing box and store it in a dry place.
The gage without the box is not allowed to be placed horzontally, nor can it be leaning against
other goods. Move the beam to the lowest positie on after use. The gage should be
calibrated regularly.