To measure the apparatus’ insulation resistance, proceed in accordance with the user
manual of REVEX 2051. Detailed measurement procedure can be found in chapter
4.3.1 on page 22. During the measurement, follow the instructions of the trainer and
always proceed only in accordance with the manual of REVEX 2051. Record the
result into the PSTC protocol and compare with the recommended maximum value.
Confirm whether the machine passed.
To measure the apparatus’ protective conductor resistance, proceed in accordance with
the user manual of REVEX 2051. Detailed measurement procedure can be found in
chapter 4.1.1 on page 13. During the measurement, follow the instructions of the
trainer and always proceed only in accordance with the manual of REVEX 2051.
Record the result into the PSTC protocol and compare with the recommended
maximum value. Confirm whether the machine passed.
To measure the leakage current via the accessible part, proceed in accordance with the
user manual of REVEX 2051. Detailed measurement procedure can be found in
chapter 4.4.1 on page 26. During the measurement, follow the instructions of the
trainer and always proceed only in accordance with the manual of REVEX 2051.
Record the result into the PSTC protocol and compare with the recommended
maximum value. Confirm whether the machine passed.
To measure the leakage current via the applied part, it is necessary to use the
measurement device RM 2050, which will be connected into REVEX 2051. Detailed
measurement procedure can be found in chapter 4.1 on page 4. During the
measurement, follow the instructions of the trainer and always proceed only in
accordance with the manual of REVEX 2051 and RM 2050. Record the result into the
PSTC protocol and compare with the recommended maximum value. Confirm
whether the machine passed.
5)
Verify the apparatus’ functional parameters according to the entries of the particular
PSTC protocol.
To verify the function of the anaesthesiological monitor, it is necessary to simulate the
electrophysiological processes in the human body. To simulate the ECG signal, a
simulator FLUKE medSim 300B will be used, or another one that is capable of
reliably simulating the heart’s electrical signals.
Connect the anaesthesiological monitor in the 5-lead ECG scanning mode to the
medSim 300B simulator. While connecting, be careful about correct contacts of the
electrodes’ clamps with the simulator; proceed in accordance with the recommended
101