User manual
connection according to the user manual for medSim 300B. Turn the monitor on and
set up the ECG measurement on at least two leads (e.g. II and V1), the pulse rate
reading and the breathing frequency measurement. Always set up the monitor in
accordance with the user manual.
According to the user manual for medSim 300B, set up several possible pulse rates
and breathing frequencies, always comparing them with the value numerically
displayed on the anaesthesiological monitor, and then check whether the ECG signal’s
course is clean, with not artefacts or distinct murmurs. When setting up the ECG
simulator, always use ECG simulation without pathological changes (ECG of a
healthy person). Record the result into the PSTC protocol and confirm whether the
machine passed.
Measuredresults
Record the results in the individual protocols. Be particular about the level of the
records, as in reality, crossed out PSTC protocol is legislatively non-permissible.
Check‐upquestionsregardingthegivenissue
• Describe the principle of the ECG scanning.
• What is the purpose of measuring the patient’s blood pressure?
• What is pulse oxymetry?
• What is the significance of STC?
Referencestousedandrecommendedinformationsources
[8.1] Wesbter, J.G. ed. Encyclopedia of Medical Devices and Instrumentation. Wiley.
[online]. c1999-2009, poslední aktualizace 17. 8. 2008 [cit. 2009-05-12]. Dostupné z
WWW: http://mrw.interscience.wiley.com/emrw/9780471732877/home/
[8.2] Rozman, J. a kol. Elektronické přístroje v lékařství. Praha: Academia, 2006. 406 s.
[8.3] Zákon č. 123/2000 Sb., zdroj: SBÍRKA ZÁKONŮ ročník 2000, částka 36, ze dne
12.5.2000.
[8.4] ČSN EN 60601-1-2. Zdravotnické elektrické přístroje: Všeobecné požadavky na
bezpečnost. 02/2003.
[8.5] ČSN EN 62353. Zdravotnické elektrické přístroje: Opakované zkoušky a zkoušky po
opravách zdravotnických elektrických přístrojů. 8/2008
102