User manual

barrier is specified by the size of the isolation voltage u
iso
, the isolation capacity C, and the
isolation resistance R. From the point of view of the medical devices and instrumentation, it is
very important that the values of the leaking currents past the isolation barrier remain as low
as possible. Low values of the direct-current leaking currents are achieved via high isolation
resistance. On the contrary, a very low value of the isolation capacity ensures low alternating
leaking currents.
The Isolation Mode Rejection Ratio (IMRR) is defined for the isolation amplifiers.
This factor indicates the isolation amplifier’s capability to reduce the disturbing common
mode voltage.
The isolation bond can be realized in several ways:
optocouplers
induction-transformer bond
capacitive bond
DC/DC converters
analogue isolation amplifiers
digital isolation devices
Secure supply circuits for ECG apparatus
From the point of view of the patient’s safety, the supply circuits must be galvanically
separated from the mains; see the technical standard ČSN EN 60601-1 and other derived
group standards.
To ensure safety and to meet the standard’s requirements, a battery supply may be
used or the DC/DC converters, which ensure the separation of the electrical circuit’s supply
from the mains and at the same time, they change the voltage size.
Additional circuits: Checking the quality of the electrodes’ connection
To sense the biopotentials via non-invasive methods, it is necessary to ensure
impedance modification of the electrode-patient transition. A special gel with chloride ions is
used for a good contact between the patient’s skin and the electrode. The impedance value of
this transition varies between single units and tens of k. The value of this impedance is very
important for correct sensing of the biopotentials, and therefore it must be checked
29