Fig. 3.5: Equivalent electrical diagram of a final stage of a monophasic defibrillator with
damp-shaped defibrillation pulse with delay. Position 1 - charging of the capacitors, position
2 - discharging of the capacitors past the delay line (defibrillation pulse DMT).
Synchronizedcardioversion
Synchronized cardioversion, or simply a cardioversion, is basically a synchronized
defibrillation, which uses lower energy values of the defibrillation pulse. Typically, the
synchronized cardioversion starts at the discharge value of 50 J in external biphasic
cardioversion and at 5 J in internal biphasic cardioversion. The defibrillator senses the ECG
(see Chapter 2) signal and calculates the distance of the R waves (the so called R-R interval),
which means the distance of subsequent QRS complexes. Cardioversion is most often used in
case of atrial flutter, atrial fibrillation and ventricular tachycardia. The applied discharge
depolarizes the entire heart muscle, thus interrupting the so called reentry circuit.
Transcutaneousstimulation(pacing)
At present, most external defibrillators can perform the so called transcutaneous
(external) stimulation besides defibrillation (TEPs - Transcutaneous External Cardiac Pacing).
Transcutaneous stimulation is used in acute arrhythmias, in case of some forms of
bradycardias and tachycardia (more than 90 beats/min, adult patients). The stimulation may
be carried out in the on-demand mode or in the fix-rate mode. Single-use electrodes are used
for the stimulation, with large surface area, the so called “pads”. The active surface of
electrodes for transcutaneous stimulation must not be less than 75 cm
2
for adult patients and
20 cm
2
for child patients.
In the on-demand mode (bradycardia, asystole), the stimulator is inactive as long as
the patient has his own cardiac activity with frequency higher than that set up on the simulator
(mostly 30-180 beats/min). The stimulation pulses are only applied in case the heart
frequency decreases or if the heart stops beating. The heart’s activity is monitored via the
36