User manual
Measurementtasks
Prior to the actual measurement, get familiar with the manipulation with and the
components of the available defibrillators (such as CardioServ, GE Healthcare [3.8] and
BeneHeart D3, Mindray [3.9]) and the defibrillator analyser (DA-2006, BC Biomedical [3.7]).
During the measurements, consult all your procedures with the teacher. Work with the user
manuals [3.8] and [3.9] available at the subject’s website.
1) With the defibrillator analyzer, compare the set and the measured parameters of the
defibrillation pulse for more set ups within the range of the apparatus (at least for 5
values of energy). Compare the defibrillation pulses in different devices and make a
chart of the measured courses.
2) Measure the charging time of the defibrillator’s capacitors.
3) Set the defibrillator analyzer to the “ECG simulator” mode, and watch different pre-
programmed cardiac arrhythmias on the defibrillator’s display. Further on, switch the
analyzer into the “Cardioversion test” mode. Set up the synchronized cardioversion
mode (discharge synchronized with the closest R-wave) according to the defibrillator’s
user manual, and measure the parameters of the synchronized discharge via the
analyzer.
4) Measure the parameters and the course of the external defibrillator pulse in the
transcutaneous stimulation mode, connected to an oscilloscope via a edutools.
Compare the measured pulse parameters with the pulse parameters from sections 2)
and 3) of the measurement. For different repetitive pulse frequencies, verify the
consistency of parameters of the transcutaneous stimulation pulses for one set up of
the stimulation current size.
5) Use the electro-revisional apparatus (MEDITEST 50) to measure the leakage current
in the applied parts. Make a protocolar record on the measurements carried out and on
the safety-technical check of the defibrillation apparatus.
Measurementprocedure
During the measurement, be careful not to interrupt the contact between the
defibrillator electrodes and the analyzer electrodes when applying the discharge, as this
would induce a spark and damage (burn) the surface of the electrodes. Also be careful
not to have the defibrillator electrodes touching at the application of the discharge.
39