Ad task 1)
Switch on the analyzer using the switch on the back side, see Fig. 3.7. Set up the
analyzer to the required range according to the energies you selected on the defibrillator, see
Fig. 3.8. Set up the scope of energy by means of the “Range” button to “High Defibrillator
Range” for discharge energy up to 1000 J, or to “Low Defibrillator Range” for discharge
energy up to 50 J. Wait until the analyzer’s display shows “Status: Ready for Defib“, and now
the analyzer is ready for the measurement.
Fig. 3.7: Defibrillator analyzer - view of the back side with legend [3.6].
On the defibrillator, choose the value of the discharge energy corresponding to the
selected range on the analyzer.
Smear the defibrillator electrodes with defibrillator gel or ECG gel (so called
Paddle) to prevent the interface surfaces from burning!!!
Place the defibrillator electrodes perfectly on the analyzer electrodes and push the
“Charge” button. At the moment the sound signalizes that the defibrillator is charged, placidly
push both the “Defib” (“Discharge”) buttons on the electrodes with your thumbs.
40