User manual
Fig. 3.8: Defibrillator analyzer - overall view with legend [3.6].
Read and record the applied discharge’s energy on the analyzer’s display - “Energy”,
the peak voltage value - “Peak V”, the peak current value - “Peak I”, and the length of the
defibrillation pulse. Use the “Playback last pulse” button to switch to the chart of the
defibrillation pulse’s duration and sketch its shape.
Ad task 2)
Use the “Range” button to switch the analyzer into the “High Defibrillator Range”
mode. Set the defibrillator to a maximum value of the discharge energy. Push the “Start
charge timer” button on the analyzer, and during the “Charge Timer Will Begin in”
countdown on the display, apply the defibrillator electrodes on the analyzer electrodes. At the
moment the countdown reaches zero, push the button for charging the defibrillator and allow
for the charging. When the charging is finished, discharge the defibrillator by means of the
41