User manual
Ad task 5)
Safety-technical check: on the basis of the operation manual [3.8, 3.9] for the
defibrillator and the electro-revisional apparatus Meditest 50 [3.7], first determine the
defibrillator’s electrical insulation class, and then perform the electrical safety measurement,
which is bound to the assigned MD class. Elaborate a protocol on STC, which you can find on
the subject’s web site.
Fig. 3.9: Converter for pacing, demonstration of stimulator and oscilloscope connection.
Apparatusandaidsused
Monophasic defibrillator CardioServ (GE Healthcare, USA)
Biphasic defibrillator BeneHeart D3 (Mindray, USA)
Defibrillator analyser DA-2006 (BC Biomedical, USA)
Converter for pacing (FBMI ČVUT in Prague)
Digital apparatus for checking medical electrical apparatus MEDITEST 50 (ILLKO,
s.r.o., Czech Rep.)
Digital oscilloscope and connecting BNC cable
Measuredresults
Elaborate a protocol on the measurement from the laboratory task. Record the
measured and calculated values into tables and draw the courses of the stimulation and
defibrillation impulses into charts with descriptions of the axes and a corresponding scale.
Conclusion
In the measurement protocol, shortly comment on each point of the measurement.
Evaluate the measured results and compare them with the data stated by the manufacturer in
43