User manual

SEV LITOVEL 2P
POLYMED ID 20/50
GOSSEN METRAWATT SECULIFE-IF + ACCESSORIES
ILLKO REVEX 2051 + ACCESSORIES
LUER syringe - 20 ml, 50 ml
500ml beaker
Infusion set
ProcedureofthePSTCelaboration:
1) Carry out a detailed inspection of the apparatus for mechanical damage.
2)
Check the integrity of the supply cable.
3)
Check the functioning of the signalizing and controlling elements.
4)
A - Measure the value of the leakage current of the apparatus and the resistance of the
protective grounding (for ARCOMED Siramed
µSP6000 and Volumed µVP7000).
B - Measure the value of the insulation resistance of the apparatus and the leakage
current via the cover of the apparatus (for SEV LITOVEL 2P and POLYMED ID
20/50).
5)
Carry out the verification of the apparatus’ functional parameters, according to the
items of the particular PSTC protocol.
For the above mentioned apparatus in points (1 to 4) carry out the tasks (1-5) and
record the results in the protocol, confirm meeting or not meeting the parameters declared in
the protocol.
Measurementprocedure
1) Carry out a detailed inspection of the apparatus for mechanical damage.
Check the apparatus very carefully to detect cracks; check namely the plastic moving
mechanisms, which may show mechanical damage more frequently. Verify the function of the
apparatus’ mechanical parts; the moving components must move freely without any marked
resistance, yet should they have an arrest in the end position, then they must not be released
from this position without exerting reasonable strength. Record the findings into the PSTC
protocol by means of checking the cell “Satisfactory“.
63